Oxygenation-ozonation of blood during extracorporeal circulation: In vitro efficiency of a new gas exchange device

被引:8
作者
Bocci, Velio
Zanardi, Iacopo
Travagli, Valter
Di Paolo, Nicola
机构
[1] Univ Siena, Dept Physiol, I-53100 Siena, Italy
[2] Univ Siena, Dept Pharmaceut Chem & Technol, I-53100 Siena, Italy
[3] Azienda Osped Sense, Dialysis & Nephrol Unit, Siena, Italy
关键词
ozone; extracorporeal circulation; polipropylene hollow fibers; phosphorylcholine coating; gas exchange device;
D O I
10.1111/j.1525-1594.2007.00448.x
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
We have investigated the performance of a new gas exchange device (GED), named L001, specifically devised for the ozonation of human blood during extracorporeal circulation. This procedure, defined with the acronym '' EBOO,'' means '' extracorporeal blood oxygenation-ozonation.'' The innovative GED is made of microporous, ozone-resistant, polipropylene hollow fibers with an external diameter of 200 mu m, a thickness of 50 mu m, and a membrane surface area of 0.22 m(2). The material is coated with phosphorylcholine on the external side in contact with the circulating blood, while a gas mixture, necessarily composed of medical oxygen and ozone (about 99 and 1%, respectively), flows inside the fibers in opposite direction. The new GED has been tested by using a buffered saline solution containing KI and by varying several parameters, and it has shown to be very versatile and efficient. Its main characteristics are minimal foreign surface contact, high gas transfer, and negligible priming volume. This device appears to be a practical, nontoxic, and rather inexpensive tool for performing ozonation of blood for already defined human diseases.
引用
收藏
页码:743 / 748
页数:6
相关论文
共 12 条
[1]   Oxygenation-ozonization of blood during extracorporeal circulation (EBOO). Part III: A new medical approach [J].
Bocci, V ;
Di Paolo, N .
OZONE-SCIENCE & ENGINEERING, 2004, 26 (02) :195-205
[2]  
Bocci V, 1999, PLATELETS, V10, P110
[3]   Ozonation of blood during extracorporeal circulation II. Comparative analysis of several oxygenator-ozonators and selection of one type [J].
Bocci, V ;
Di Paolo, N ;
Borrelli, E ;
Larini, A ;
Cappelletti, F .
INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 2001, 24 (12) :890-897
[4]   Ozonation of blood during extracorporeal circulation. I. Rationale, methodology and preliminary studies [J].
Bocci, V ;
Di Paolo, N ;
Garosi, G ;
Aldinucci, C ;
Borrelli, E ;
Valacchi, G ;
Cappelli, F ;
Guerri, L ;
Gavioli, G ;
Corradeschi, F ;
Rossi, R ;
Giannerini, F ;
Di Simplicio, P .
INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 1999, 22 (09) :645-651
[5]  
BOCCI V, 2005, NEW MED DRUG
[6]   Biochemical modifications induced in human blood by oxygenation-ozonation [J].
Bocci, Velio ;
Aldinucci, Carlo .
JOURNAL OF BIOCHEMICAL AND MOLECULAR TOXICOLOGY, 2006, 20 (03) :133-138
[7]   Phosphorylcholine coating of extracorporeal circuits provides natural protection against blood activation by the material surface [J].
De Somer, F ;
François, K ;
van Oeveren, W ;
Poelaert, J ;
De Wolf, D ;
Ebels, T ;
Van Nooten, G .
EUROPEAN JOURNAL OF CARDIO-THORACIC SURGERY, 2000, 18 (05) :602-606
[8]   Extracorporeal blood oxygenation and ozonation (EBOO): A controlled trial in patients with peripheral artery disease [J].
Di Paolo, N ;
Bocci, V ;
Salvo, DP ;
Palasciano, F ;
Biagioli, M ;
Meini, S ;
Galli, F ;
Ciari, I ;
Maccari, F ;
Cappelletti, F ;
Di Paolo, M ;
Gaggiotti, E .
INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 2005, 28 (10) :1039-1050
[9]   Extracorporeal blood oxygenation and ozonation (EBOO) in man. Preliminary report [J].
Di Paolo, N ;
Bocci, V ;
Garosi, G ;
Borrelli, E ;
Bravi, A ;
Bruci, A ;
Aldinucci, C ;
Capotondo, L .
INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 2000, 23 (02) :131-141
[10]   Necrotizing fasciitis successfully treated with extracorporeal blood oxygenation and ozonization (EBOO) [J].
Di Paolo, N ;
Bocci, V ;
Cappelletti, F ;
Petrini, G ;
Gaggiotti, E .
INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 2002, 25 (12) :1194-1198