In vitro performance characteristics of high-frequency oscillatory ventilators

被引:61
作者
Pillow, JJ
Wilkinson, MH
Neil, HL
Ramsden, CA
机构
[1] TVW Telethon Inst Child Hlth Res, Perth, WA, Australia
[2] Univ Western Australia, Dept Paediat, Perth, WA 6009, Australia
[3] King Edqard & Princess Margaret Hosp, Perth, WA, Australia
[4] Monash Univ, Inst Reprod & Dev, Ritchie Ctr Baby Hlth Res, Clayton, Vic 3168, Australia
[5] Monash Univ, Dept Paediat, Clayton, Vic 3168, Australia
关键词
compliance; high-frequency ventilation; newborn; respiratory distress syndrome; ventilator properties;
D O I
10.1164/ajrccm.164.6.2005008
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
This study aimed to examine the performance characteristics of four high-frequency oscillatory-type ventilators, using an in vitro model of the intubated neonatal respiratory system. Each ventilator was examined across its operative range of settings and at varying model lung compliance (C) and resistance. The oscillatory pressure waveform was measured at the airway opening (Pao). Tidal volume (VT) and flow were determined from pressure changes within the model lung (Delta PA). The spectral content of the Pao waveform differed between ventilators. The maximum ventilator VT ranged from 3.7 to 11.1 ml at 15 Hz and a mean airway pressure (Paw) of 12 cm H2O to oscillate a model lung (C = 0.4 ml/cm H2O) through a 3.0-mm internal diameter (i.d.) endotracheal tube (ETT). A small drop in C was associated with a decrease in VT and marked increase in Delta PA from 0.1 to 0.8 ml/cm H2O. The influence of C on VT and Delta PA and the pressure cost of ventilation (Delta PA/f.VT2) was dependent on the oscillatory frequency, ETT inner diameter, and the specific ventilator used. Substantive differences exist between oscillatory ventilators that need to be considered in their clinical application. The rapid establishment of optimal lung volume and oscillatory frequency is important in minimizing barotrauma during high-frequency oscillatory ventilation.
引用
收藏
页码:1019 / 1024
页数:6
相关论文
共 27 条
[1]  
CLARK RH, 1992, PEDIATRICS, V89, P5
[2]   RESPIRATORY SYSTEM IMPEDANCE FROM 4 TO 40 HZ IN PARALYZED INTUBATED INFANTS WITH RESPIRATORY-DISEASE [J].
DORKIN, HL ;
STARK, AR ;
WERTHAMMER, JW ;
STRIEDER, DJ ;
FREDBERG, JJ ;
FRANTZ, ID .
JOURNAL OF CLINICAL INVESTIGATION, 1983, 72 (03) :903-910
[3]  
DURAND DJ, 1993, J PEDIATR-US, V122, P609
[4]   ALVEOLAR PRESSURE SWINGS DURING HIGH-FREQUENCY VENTILATION IN RABBITS [J].
FRANTZ, ID ;
CLOSE, RH .
PEDIATRIC RESEARCH, 1985, 19 (02) :162-166
[5]   AUGMENTED DIFFUSION IN THE AIRWAYS CAN SUPPORT PULMONARY GAS-EXCHANGE [J].
FREDBERG, JJ .
JOURNAL OF APPLIED PHYSIOLOGY, 1980, 49 (02) :232-238
[6]   FACTORS INFLUENCING MECHANICAL PERFORMANCE OF NEONATAL HIGH-FREQUENCY VENTILATORS [J].
FREDBERG, JJ ;
GLASS, GM ;
BOYNTON, BR ;
FRANTZ, ID .
JOURNAL OF APPLIED PHYSIOLOGY, 1987, 62 (06) :2485-2490
[7]   ROLE OF LUNG-VOLUME IN LUNG INJURY - HFO IN THE ATELECTASIS-PRONE LUNG [J].
FROESE, AB .
ACTA ANAESTHESIOLOGICA SCANDINAVICA, 1989, 33 :126-130
[8]  
FROESE AB, 1987, ANESTH ANALG, V66, P814
[9]   PRESSURE-FLOW RELATIONSHIPS OF ENDOTRACHEAL-TUBES DURING HIGH-FREQUENCY VENTILATION [J].
GAVRIELY, N ;
SOLWAY, J ;
LORING, SH ;
BUTLER, JP ;
SLUTSKY, AS ;
DRAZEN, JM .
JOURNAL OF APPLIED PHYSIOLOGY, 1985, 59 (01) :3-11
[10]  
Gerstmann DR, 1996, PEDIATRICS, V98, P1044