Radiation doses in interventional radiology procedures: The RAD-IR study Part III: Dosimetric performance of the interventional fluoroscopy units

被引:45
作者
Balter, S
Schueler, BA
Miller, DL [1 ]
Cole, PE
Lu, HT
Berenstein, A
Albert, R
Georgia, JD
Noonan, PT
Russell, EJ
Malisch, TW
Vogelzang, RL
Geisinger, M
Cardella, JF
St George, J
Miller, GL
Anderson, J
机构
[1] Natl Naval Med Ctr, Dept Radiol, Bethesda, MD 20889 USA
[2] Lenox Hill Hosp, Dept Med, New York, NY 10021 USA
[3] Mayo Clin, Rochester, MN USA
[4] Uniformed Serv Univ Hlth Sci, F Edward Hebert Sch Med, Dept Radiol & Radiol Sci, Bethesda, MD 20814 USA
[5] NCI, Med Oncol Clin Res Unit, Canc Res Ctr, Bethesda, MD 20892 USA
[6] US FDA, Ctr Devices & Radiol Hlth, Rockville, MD 20857 USA
[7] Beth Israel Deaconess Med Ctr, Ctr Endovasc Surg, Hyman Newman Inst Neurol & Neurosurg, New York, NY 10003 USA
[8] NW Mem Hosp, Dept Radiol, Chicago, IL 60611 USA
[9] Northwestern Univ, Feinberg Sch Med, Dept Radiol, Chicago, IL 60611 USA
[10] Cleveland Clin Fdn, Dept Radiol, Cleveland, OH 44195 USA
[11] SUNY Upstate Med Univ, Dept Radiol, Syracuse, NY USA
[12] Univ Texas, SW Med Ctr, Dept Radiol, Dallas, TX 75235 USA
[13] Lahey Clin Fdn, Dept Radiol, Burlington, MA USA
[14] Univ Colorado, Hlth Sci Ctr, Dept Radiol, Denver, CO 80262 USA
关键词
D O I
10.1097/01.RVI.0000130864.68139.08
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
PURPOSE: To present the physics data supporting the validity of the clinical dose data from the RAD-IR study and to document the performance of dosimetry-components of these systems over time. MATERIALS AND METHODS: Sites at seven academic medical centers in the United States prospectively contributed data for each of 12 fluoroscopic units. All units were compatible with International Electrotechnical Commission (IEC) standard 60601-2-43. Comprehensive evaluations and periodic consistency checks were performed to verify the performance of each unit's dosimeter. Comprehensive evaluations compared system performance against calibrated ionization chambers under nine combinations of operating conditions. Consistency checks provided more frequent dosimetry data, with use of each unit's built-in dosimetry equipment and a standard water phantom. RESULTS: During the 3-year study, data were collected for 48 comprehensive evaluations and 581 consistency checks. For the comprehensive evaluations, the mean (95% confidence interval range) ratio of system to external measurements was 1.03 (1.00-1.05) for fluoroscopy and 0.93 (0.90-0.96) for acquisition. The expected ratio was 0.93 for both. For consistency checks, the values were 1.00 (0.98-1.02) for fluoroscopy and 1.00 (0.98-1.02) for acquisition. Each system was compared across time to its own mean value. Overall uncertainty was estimated by adding the standard deviations of the comprehensive and consistency measurements in quadrature. The authors estimate that the overall error in clinical cumulative dose measurements reported in RAD-IR is 24%. CONCLUSION: Dosimetric accuracy was well within the tolerances established by IEC standard 60601-2-43. The clinical dose data reported in the RAD-IR study are valid.
引用
收藏
页码:919 / 926
页数:8
相关论文
共 9 条
[1]  
[Anonymous], 2000, 60601 IEC
[2]  
GEISER WR, 1997, MED PHYS, V24, P337
[3]  
International Electrotechnical Commission, 2000, 60580 IEC
[4]   Radiation protection in interventional radiology [J].
Johnson, DR ;
Kyriou, J ;
Morton, EJ ;
Clifton, A ;
Fitzgerald, M ;
MacSweeney, E .
CLINICAL RADIOLOGY, 2001, 56 (02) :99-106
[5]   Radiation doses in interventional radiology procedures: The RAD-IR Study - Part I: Overall measures of dose [J].
Miller, DL ;
Balter, S ;
Cole, PE ;
Lu, HT ;
Schueler, BA ;
Geisinger, M ;
Berenstein, A ;
Albert, R ;
Georgia, JD ;
Noonan, PT ;
Cardella, JF ;
George, JS ;
Russell, EJ ;
Malisch, TW ;
Vogelzang, RL ;
Miller, GL ;
Anderson, J .
JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2003, 14 (06) :711-727
[6]  
Miller DL, 2003, J VASC INTERV RADIOL, V14, P977
[7]   AN EVALUATION OF THE DIAMENTOR TRANSMISSION IONIZATION-CHAMBER IN INDICATING EXPOSURE AREA PRODUCT (R-CM2) DURING DIAGNOSTIC RADIOLOGICAL EXAMINATIONS [J].
SHRIMPTON, PC ;
WALL, BF .
PHYSICS IN MEDICINE AND BIOLOGY, 1982, 27 (06) :871-878
[8]  
1994, FED REG, V59, P26402
[9]  
2002, FED REG, V67, P76056