Fast 3D large-angle spin-echo imaging (3D FLASE)

被引:84
作者
Ma, JF [1 ]
Wehrli, FW [1 ]
Song, HK [1 ]
机构
[1] UNIV PENN,MED CTR,DEPT RADIOL,PHILADELPHIA,PA 19104
关键词
large flip angle imaging; spin echo; microscopy; trabecular bone;
D O I
10.1002/mrm.1910350619
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
A rapid steady-state 3D spin-echo imaging pulse sequence, based on the principle of nutating the spins by an angle greater than 90 degrees, has been designed and implemented on a clinical 1.5-T whole-body MR scanner. The pulse sequence, denoted fast large-angle spin-echo (FLASE), has been optimized for high-resolution imaging of tissues with short T-2 and T-2*. Features of FLASE include a minimum-phase Shinnar-Le Roux excitation pulse and distribution of phase- and slice-encoding gradients before and after the 180 degrees refocusing pulse to minimize the critical time delay between inversion and restoration of the residual longitudinal magnetization and for minimizing echo time, A Bloch equation analysis, corroborated by experimental data, shows FLASE signal-to-noise to be superior to its closest analog, 3D rapid spin-echo excitation (RASEE) (Jara et al., Magn Reson Medicine 29,528 (1993)), and 3D gradient-recalled acquisition in steady state (GRASS). It is demonstrated that with judicious RF phase-cycling and steady state operation, FLASE can produce high-quality microimages free of intravoxel phase dispersion from susceptibility-induced background gradients. The performance of the method is exemplified with ultra high-resolution images of trabecular bone in vitro and in vivo in the human calcaneus and wrist at voxel sizes as low as 98 x 98 x 200 mu m(3). Finally, the contrast behavior of refocused FLASE can be altered by disrupting the steady state analogous to gradient echo imaging.
引用
收藏
页码:903 / 910
页数:8
相关论文
共 32 条
[1]   RASEE - A RAPID SPIN-ECHO PULSE SEQUENCE [J].
BOGDAN, AR ;
JOSEPH, PM .
MAGNETIC RESONANCE IMAGING, 1990, 8 (01) :13-19
[2]   CONTRAST IN RAPID MR IMAGING - T1-WEIGHTED AND T2-WEIGHTED IMAGING [J].
BUXTON, RB ;
EDELMAN, RR ;
ROSEN, BR ;
WISMER, GL ;
BRADY, TJ .
JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1987, 11 (01) :7-16
[3]   NUCLEAR MAGNETIC-RESONANCE MICROSCOPY WITH 4-MU-M RESOLUTION - THEORETICAL-STUDY AND EXPERIMENTAL RESULTS [J].
CHO, ZH ;
AHN, CB ;
JUH, SC ;
LEE, HK ;
JACOBS, RE ;
LEE, S ;
YI, JH ;
JO, JM .
MEDICAL PHYSICS, 1988, 15 (06) :815-824
[4]  
CHUN GH, 1995, J BONE MINER RES, V10, P803
[5]   THE LOSS OF SMALL OBJECTS IN VARIABLE TE IMAGING - IMPLICATIONS FOR FSE, RARE, AND EPI [J].
CONSTABLE, RT ;
GORE, JC .
MAGNETIC RESONANCE IN MEDICINE, 1992, 28 (01) :9-24
[6]   LARGE-ANGLE SPIN-ECHO IMAGING [J].
DIIORIO, G ;
BROWN, JJ ;
BORRELLO, JA ;
PERMAN, WH ;
SHU, HH .
MAGNETIC RESONANCE IMAGING, 1995, 13 (01) :39-44
[7]   APPLICATION OF FOURIER TRANSFORM SPECTROSCOPY TO MAGNETIC RESONANCE [J].
ERNST, RR ;
ANDERSON, WA .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1966, 37 (01) :93-+
[8]   HIGH-RESOLUTION MR IMAGING OF THE WRIST AND EYE WITH SHORT TR, SHORT TE, AND PARTIAL-ECHO ACQUISITION [J].
FOO, TKF ;
SHELLOCK, FG ;
HAYES, CE ;
SCHENCK, JF ;
SLAYMAN, BE .
RADIOLOGY, 1992, 183 (01) :277-281
[9]   DIRECT FLASH MR IMAGING OF MAGNETIC-FIELD INHOMOGENEITIES BY GRADIENT COMPENSATION [J].
FRAHM, J ;
MERBOLDT, KD ;
HANICKE, W .
MAGNETIC RESONANCE IN MEDICINE, 1988, 6 (04) :474-480
[10]  
FRAHM J, 1987, J MAGN RESON, V27, P307