Diagnostic assessment of recurrent unexplained syncope with a new subcutaneously implantable loop recorder

被引:51
作者
Seidl, K
Rameken, M
Breunung, S
Senges, J
Jung, W
Andresen, D
van Toor, A
Krahn, AD
Klein, GJ
机构
[1] Herzzentrum Ludwigshafen, D-67063 Ludwigshafen, Germany
[2] Med Univ Klin & Poliklin, Bonn, Germany
[3] Urban Krankenhaus, Med Klin Kardiol 1, Berlin, Germany
[4] Bakken Res Ctr BV, Maastricht, Netherlands
[5] Univ Western Ontario, Univ Hosp, Dept Med, Div Cardiol, London, ON N6A 5A5, Canada
来源
EUROPACE | 2000年 / 2卷 / 03期
关键词
syncope; implantable device; loop recorder;
D O I
10.1053/eupc.2000.0108
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Aim Patients with recurrent syncope undiagnosed after extensive non-invasive and invasive testing pose a diagnostic and therapeutic dilemma. Holter monitoring is nondiagnostic in 90% of cases. Recent developments in loop recorder technology permit long-term ECG monitoring in patients with recurrent unexplained syncope. The aim of this study was to report the worldwide experience with a new subcutaneously implantable loop recorder, implanted in 133 patients with unexplained syncope and negative laboratory investigations. Methods and Results The implantable loop recorder monitors continuously a single lead electrogram using two sensing electrodes on the device shell. The device was implanted in 133 patients, 67 male and 66 female with recurrent syncope. During a mean follow-up of 10.8 +/- 4.3 months after device implantation, 83 patients (62%) experienced syncope or pre-syncope. In the remaining 50 patients no diagnosis could be made because either no events occurred, the patients were lost to follow-up, had adverse events, or died prior to diagnosis. In 72 of the 83 patients with syncope during follow-up (87%), loop recording definitively determined whether an arrhythmia was the cause of symptoms or not. Diagnosis included bradycardia in 21 patients, pacemaker dysfunction in one patient, and tachycardia in 10 patients. One patient experienced multiple rhythm disturbances. Syncope was non-arrhythmic in 40 patients. The remaining 11 patients failed to press the activator. Therapy was instituted in all patients, in whom an arrhythmic cause was found. Severe anticipated device related complications occurred in three patients. Conclusion An implantable loop recorder is useful for establishing a diagnosis when symptoms are recurrent but too infrequent for conventional monitoring techniques. (Europace 2000; 2: 256-262) (C) 2000 The European Society of Cardiology.
引用
收藏
页码:256 / 262
页数:7
相关论文
共 25 条
[1]  
AKTHAR M, 1983, PACE, V6, P192
[2]   UTILITY OF ELECTROPHYSIOLOGIC STUDIES IN THE MANAGEMENT OF TACHYCARDIA, SUDDEN-DEATH, AND SYNCOPE [J].
BARDY, GH ;
PACKER, DL ;
GERMAN, LD ;
GALLAGHER, JJ .
ANNALS OF THE NEW YORK ACADEMY OF SCIENCES, 1984, 427 (JUN) :16-39
[3]   LONG-TERM PROGNOSIS OF PATIENTS UNDERGOING ELECTROPHYSIOLOGIC STUDIES FOR SYNCOPE OF UNKNOWN ORIGIN [J].
BASS, EB ;
ELSON, JJ ;
FOGOROS, RN ;
PETERSON, J ;
ARENA, VC ;
KAPOOR, WN .
AMERICAN JOURNAL OF CARDIOLOGY, 1988, 62 (17) :1186-1191
[4]  
BROWN AP, 1987, BRIT HEART J, V58, P251
[5]   ARRHYTHMIAS DETECTED BY AMBULATORY MONITORING - LACK OF CORRELATION WITH SYMPTOMS OF DIZZINESS AND SYNCOPE [J].
CLARK, PI ;
GLASSER, SP ;
SPOTO, E .
CHEST, 1980, 77 (06) :722-725
[6]   EVALUATION AND OUTCOME OF EMERGENCY ROOM PATIENTS WITH TRANSIENT LOSS OF CONSCIOUSNESS [J].
DAY, SC ;
COOK, EF ;
FUNKENSTEIN, H ;
GOLDMAN, L .
AMERICAN JOURNAL OF MEDICINE, 1982, 73 (01) :15-23
[7]  
DENNISS AR, 1992, INT J CARDIOL, V35, P211
[8]  
DIMARCO JP, 1990, ANN INTERN MED, V113, P53, DOI 10.7326/0003-4819-113-1-53
[9]   DIAGNOSTIC EFFICACY OF 24-HOUR ELECTROCARDIOGRAPHIC MONITORING FOR SYNCOPE [J].
GIBSON, TC ;
HEITZMAN, MR .
AMERICAN JOURNAL OF CARDIOLOGY, 1984, 53 (08) :1013-1017
[10]   VALUE AND LIMITATIONS OF CLINICAL ELECTROPHYSIOLOGIC STUDY IN ASSESSMENT OF PATIENTS WITH UNEXPLAINED SYNCOPE [J].
GULAMHUSEIN, S ;
NACCARELLI, GV ;
KO, PT ;
PRYSTOWSKY, EN ;
ZIPES, DP ;
BARNETT, HJM ;
HEGER, JJ ;
KLEIN, GJ .
AMERICAN JOURNAL OF MEDICINE, 1982, 73 (05) :700-705