Properties of perimetric threshold estimates from full threshold, ZEST, and SITA-like strategies, as determined by computer simulation

被引:109
作者
Turpin, A
McKendrick, AM
Johnson, CA
Vingrys, AJ
机构
[1] Univ Western Australia, Sch Psychol, Crawley, WA 6009, Australia
[2] Curtin Univ Technol, Dept Comp, Perth, WA, Australia
[3] Devers Eye Inst, Legacy Clin Res & Technol Ctr, Portland, OR USA
[4] Univ Melbourne, Dept Optometry & Vis Sci, Melbourne, Vic, Australia
关键词
D O I
10.1167/iovs.03-0023
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
PURPOSE. To investigate the accuracy and precision of threshold estimates returned by two Bayesian perimetric strategies, staircase-QUEST or SQ (a Swedish interactive threshold algorithm [SITA]-like strategy) and ZEST (zippy estimation by sequential testing), and to compare these measures with those of the full-threshold (FT) algorithm. METHODS. A computerized visual field simulation model was developed to compare the performance (accuracy, precision, and number of presentations) of the three algorithms. SQ implemented aspects of the SITA algorithm that are in the public domain. The simulation was tested by using standard automated perimetry (SAP) visual field data from 265 normal subjects and 163 observers with glaucomatous visual field loss and by exploring the effect of response variability and response errors on algorithm performance. RESULTS. SQ was faster than FT or ZEST, with a comparable mean error when simulating field tests on patients. Point-wise analysis revealed similar error and standard deviation of error as a function of threshold for FT and SQ. If the initial estimate of threshold for either procedure was incorrect, the means and standard deviations of the error increased markedly. ZEST produced more accurate thresholds than did the other two strategies when the initial estimate was removed from the true threshold. CONCLUSIONS. When simulated patients made errors, the accuracy and precision of sensitivity estimates were poor when the initial estimate of threshold either overestimated or underestimated the true threshold. This was particularly so for FT and SQ. ZEST demonstrated more consistent error properties than the other two measures.
引用
收藏
页码:4787 / 4795
页数:9
相关论文
共 23 条
[1]   Utility of a dynamic termination criterion in the ZEST adaptive threshold method [J].
Anderson, AJ .
VISION RESEARCH, 2003, 43 (02) :165-170
[2]  
Anderson D.R., 1999, Automated static perimetry, V2nd
[3]  
Artes PH, 2002, INVEST OPHTH VIS SCI, V43, P2654
[4]   SITA Fast, a new rapid perimetric threshold test. Description of methods and evaluation in patients with manifest and suspect glaucoma [J].
Bengtsson, B ;
Heijl, A .
ACTA OPHTHALMOLOGICA SCANDINAVICA, 1998, 76 (04) :431-437
[5]   Evaluation of a new perimetric threshold strategy, SITA, in patients with manifest and suspect glaucoma [J].
Bengtsson, B ;
Heijl, A .
ACTA OPHTHALMOLOGICA SCANDINAVICA, 1998, 76 (03) :268-272
[6]  
Bengtsson B, 1997, ACTA OPHTHALMOL SCAN, V75, P368
[7]   Evaluation of a new threshold visual field strategy, SITA, in normal subjects [J].
Bengtsson, B ;
Heijl, A ;
Olsson, J .
ACTA OPHTHALMOLOGICA SCANDINAVICA, 1998, 76 (02) :165-169
[8]  
Budenz DL, 2002, ARCH OPHTHALMOL-CHIC, V120, P1136
[9]  
CHAUHAN BC, 1993, INVEST OPHTH VIS SCI, V34, P3534
[10]   TEST-RETEST VARIABILITY IN GLAUCOMATOUS VISUAL-FIELDS [J].
HEIJL, A ;
LINDGREN, A ;
LINDGREN, G .
AMERICAN JOURNAL OF OPHTHALMOLOGY, 1989, 108 (02) :130-135