Amperometric micro-biosensor array for choline and acetylcholine based on a glassy carbon electrode modified by platinum-black particles and a co-crosslinked enzyme system

被引:15
作者
Chen, J [1 ]
Lin, XQ [1 ]
Chen, ZH [1 ]
Wu, SG [1 ]
Wang, SQ [1 ]
机构
[1] Univ Sci & Technol China, Dept Chem, Hefei 230026, Anhui, Peoples R China
关键词
choline; acetylcholine; amperometry; microbiosensor array; platinum-black; choline oxidase; acetylcholineasterase;
D O I
10.1081/AL-100002590
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A glassy carbon electrode was first modified by electro-depositing sub-micrometer platinum black particles on the surface. An acetylcholine micro-biosensor array and a choline micro-biosensor array were fabricated based on immobilization of acetylcholineasterase-choline oxidase or choline oxidase by cross-linking with glutaraldehyde on the Pt-black/GCE. Significant enhancement in performances of these biosensors was achieved. The chronoamperometric response of the acetylcholine micro-biosensor array gave a linear relation in the range of 29 to 1200 mu mol/L and a detection limit of 8.7 mu mol/L acetylcholine. The amperometric response of the choline micro-biosensor array gave a linear relation in the range of 24 to 1000 mu mol/L and a detection limit of 7.4 mu mol/L choline. They also possessed very good stability and reproducibility for long-term use.
引用
收藏
页码:491 / 501
页数:11
相关论文
共 19 条
[1]  
ANTONIO G, 1995, ANALYST, V120, P2731
[2]  
Chen Q, 1998, ELECTROANAL, V10, P94, DOI 10.1002/(SICI)1521-4109(199802)10:2<94::AID-ELAN94>3.0.CO
[3]  
2-A
[4]  
*COOPER JR, 1991, BIOCHCAL BASIS NEURO, P190
[5]   Covalently immobilized choline oxidase and cholinesterases on a methacrylate copolymer for disposable membrane biosensors [J].
Doretti, L ;
Gattolin, P ;
Burla, A ;
Ferrara, D ;
Lora, S ;
Palma, G .
APPLIED BIOCHEMISTRY AND BIOTECHNOLOGY, 1998, 74 (01) :1-12
[6]   Amperometric sensor for hydrogen peroxide, based on Cu2O or CuO modified carbon paste electrodes [J].
Garjonyte, R ;
Malinauskas, A .
FRESENIUS JOURNAL OF ANALYTICAL CHEMISTRY, 1998, 360 (01) :122-123
[7]   THE MECHANISM OF THE ELECTROGENERATED CHEMI-LUMINESCENCE OF LUMINOL IN AQUEOUS ALKALINE-SOLUTION [J].
HAAPAKKA, KE ;
KANKARE, JJ .
ANALYTICA CHIMICA ACTA, 1982, 138 (JUN) :263-275
[8]   ENZYME MICROELECTRODES FOR CHOLINE AND ACETYLCHOLINE AND THEIR APPLICATIONS [J].
KANO, K ;
MORIKAGE, K ;
UNO, B ;
ESAKA, Y ;
GOTO, M .
ANALYTICA CHIMICA ACTA, 1994, 299 (01) :69-74
[9]   Electrocatalytic oxidation of H2O2 at an oxycobalt film modified glassy carbon electrode for fermentation monitoring [J].
Khoo, SB ;
Yap, MGS ;
Huang, YL ;
Guo, SX .
ANALYTICA CHIMICA ACTA, 1997, 351 (1-3) :133-142
[10]   ROTATED AND STATIONARY PLATINUM WIRE ELECTRODES - RESIDUAL CURRENT-VOLTAGE CURVES AND DISSOLUTION PATTERNS IN SUPPORTING ELECTROLYTES [J].
KOLTHOFF, IM ;
TANAKA, N .
ANALYTICAL CHEMISTRY, 1954, 26 (04) :632-636