Correlation between central corneal thickness, applanation tonometry, and direct intracameral IOP readings

被引:162
作者
Feltgen, N [1 ]
Leifert, D [1 ]
Funk, J [1 ]
机构
[1] Univ Freiburg, Augenklin, Dept Ophthalmol, D-79106 Freiburg, Germany
关键词
D O I
10.1136/bjo.85.1.85
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Background-Several authors reported incorrect high intraocular pressure (IOP) values in eyes with a thick cornea using applanation tonometry. This hypothesis was checked by comparing applanation tonometry with direct intracameral manometry. Methods-73 patients, scheduled for intraocular surgery, were enrolled. Immediately before surgery, the following were registered: (i) central corneal thickness (CCT), (ii) applanatory IOP (Perkins/ Tonopen), and (iii) intracameral IOP. Results-The difference between applanatory and intraocular measurements was completely independent of CCT (y=-3.43+3.8x; where y is the difference between applanatory and intracamerally measured IOP (mm Hg) and x is CCT (mm); r(2)=0.002; p=0.72). Conclusions-There is no systematic error of applanation tonometry with increasing CCT Therefore it is inadequate to recalculate IOP based on regression formula of applanatory IOP versus CCT.
引用
收藏
页码:85 / 87
页数:3
相关论文
共 15 条
[1]   Ocular hypertension and central corneal thickness [J].
Argus, WA .
OPHTHALMOLOGY, 1995, 102 (12) :1810-1812
[2]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[3]  
Copt RP, 1999, ARCH OPHTHALMOL-CHIC, V117, P14
[4]  
EHLERS N, 1975, ACTA OPHTHALMOL, V53, P34
[5]  
GOLDMANN H, 1957, Ophthalmologica, V134, P221
[6]  
GOLDMANN H, 1961, Ophthalmologica, V141, P441
[7]  
HANSEN FK, 1971, ACTA OPHTHALMOL, V49, P775
[8]   Central corneal thickness in normal, glaucomatous, and ocular hypertensive eyes [J].
Herndon, LW ;
Choudhri, SA ;
Cox, T ;
Damji, KF ;
Shields, MB ;
Allingham, RR .
ARCHIVES OF OPHTHALMOLOGY, 1997, 115 (09) :1137-1141
[9]  
Herndon LW, 1998, ARCH OPHTHALMOL-CHIC, V116, P1263
[10]  