Evaluation of the Accusport® Lactate Analyzer

被引:71
作者
Fell, JW
Rayfield, JM
Gulbin, JP
Gaffney, PT
机构
[1] Griffith Univ, Sch Hlth Sci, NHS, Gold Coast Mail Ctr, Qld 9726, Australia
[2] Royal Brisbane Hosp, Dept Pathol, Brisbane, Qld, Australia
关键词
lactate; blood lactate; Accusport; lactate measurement; reflectance photometry;
D O I
10.1055/s-2007-971904
中图分类号
G8 [体育];
学科分类号
04 ; 0403 ;
摘要
An investigation of the Accusport(R) Lactate Analyser (ACC) was undertaken to assess its accuracy, linearity, reliability, and versatility. Accuracy was determined by comparing 32 blood assays on the ACC with those obtained using the kodak Ektachem(R) E250 (E250). Linearity was determined over the range 1.2-18.7 mmol/L. Reliability was measured at 1.7 (low) and 14.4 (high) mmol/L concentrations. Versatility involved determining the optimal sample volume, influence of haematocrit, and the effect of time delay from sampling to assay. Assays determined using the ACC confirm good agreement with the E250. The difference between the two methods compared for lactate determination does not exceed 1.1 mmol/L for concentrations up to approximately 20 mmol/L, Linearity is demonstrated by the regression equation y = 0.026 + 0.970 x, r = 0.995. Reliability is expressed by the coefficient of variation of 7% (low) and 4.6% (high). The optimal sample size for accurate assay was determined to be between 20 mu L and 50 mu L. Mean lactate value did not vary significantly between a haematocrit of 35% and 45%. The Accusport is accurate, linear up to at least 18.7 mmol/L, has good reliability at high and low concentrations, and is able to analyse whole blood up to 15 minutes after sample application to the reagent strip.
引用
收藏
页码:199 / 204
页数:6
相关论文
共 20 条
[1]  
*AM DIAB ASS, 1995, DIABETES CARE, V18, P46
[2]  
ASTLES R, 1994, CLIN CHEM, V40, P1327
[3]   BLOOD LACTATE MEASUREMENT IN RECOVERY AS AN ADJUNCT TO TRAINING - PRACTICAL CONSIDERATIONS [J].
BISHOP, P ;
MARTINO, M .
SPORTS MEDICINE, 1993, 16 (01) :5-13
[4]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[5]  
BONIFAZI M, 1993, J SPORT MED PHYS FIT, V33, P13
[6]  
Di Prampero PE, 1981, REV PHYSL BIOCH PHAR, V89, P144
[7]  
DOHERTY M, 1992, COACHING FOCUS, V21, P7
[8]   LACTATE CONCENTRATION DIFFERENCES IN PLASMA, WHOLE-BLOOD, CAPILLARY FINGER BLOOD AND ERYTHROCYTES DURING SUBMAXIMAL GRADED-EXERCISE IN HUMANS [J].
FOXDAL, P ;
SJODIN, B ;
RUDSTAM, H ;
OSTMAN, C ;
OSTMAN, B ;
HEDENSTIERNA, GC .
EUROPEAN JOURNAL OF APPLIED PHYSIOLOGY, 1990, 61 (3-4) :218-222
[9]  
GAMBKE B, 1993, 33 GERM C SPORTS MED
[10]   MEASUREMENT OF ANAEROBIC CAPACITIES IN HUMANS - DEFINITIONS, LIMITATIONS AND UNSOLVED PROBLEMS [J].
GREEN, S ;
DAWSON, B .
SPORTS MEDICINE, 1993, 15 (05) :312-327