Influence of cuff volume on oropharyngeal leak pressure and fibreoptic position with the laryngeal mask airway

被引:84
作者
Keller, C
Pühringer, F
Brimacombe, JR [1 ]
机构
[1] Univ Queensland, Cairns Base Hosp, Dept Anaesthesia & Intens Care, Cairns 4870, Australia
[2] Leopold Franzens Univ, Dept Anaesthesia & Intens Care Med, Innsbruck, Austria
关键词
equipment; masks anaesthesia; measurement techniques; fibreoptic;
D O I
10.1093/bja/81.2.186
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
We studied the size 4 laryngeal mask airway (LMA) to test the hypothesis that oropharyngeal teak pressure and fibreoptic position improves with increasing cuff volume. After LMA insertion, 50 anaesthetized adult patients had the cuff inflated in 5-ml increments to 40 mi. Oropharyngeal leak pressure was optimal at 15 mi and decreased at higher volumes. The fibreoptic position was optimal at 0-20 mt and deteriorated at higher volumes. Gastric insufflation was detected more frequently when the cuff volume exceeded 20 mi. We conclude that inflation of the size 4 LMA to the maximum recommended volume provides suboptimal conditions and that this value should be reduced from 30 to 20 ml.
引用
收藏
页码:186 / 187
页数:2
相关论文
共 5 条
[1]  
BRIMACOMBE J, 1993, ANESTH ANALG, V76, P457
[2]  
Brimacombe J, 1996, LARYNGEAL MASK AIRWA, V3rd
[3]   THE LARYNGEAL MASK AIRWAY AND POSITIVE-PRESSURE VENTILATION [J].
DEVITT, JH ;
WENSTONE, R ;
NOEL, AG ;
ODONNELL, MP .
ANESTHESIOLOGY, 1994, 80 (03) :550-555
[4]   The effect of laryngeal mask airway insertion on the position of the internal jugular vein [J].
Nandwani, N ;
Fairfield, MC ;
Krarup, K ;
Thompson, J .
ANAESTHESIA, 1997, 52 (01) :77-79
[5]   The laryngeal mask airway: A comparison between two insertion techniques [J].
Wakeling, HG ;
Butler, PJ ;
Baxter, PJC .
ANESTHESIA AND ANALGESIA, 1997, 85 (03) :687-690