摘要
The key to remediative processes is the ability to measure toxic contaminants on-site using simple and cheap sensing devices, which are field-portable and can facilitate more rapid decision-making. A three-electrode configuration system has been fabricated using low-cost screen-printing (thick-film) technology and this coupled with a portable electrochemical instrument has provided a a relatively inexpensive on-site detector for trace levels of toxic metals. The carbon surface of the screen-printed working electrode is used as a substrate for in situ deposition of a metallic film of bismuth, which allows the electrochemical preconcentration of metal ions. Lead and cadmium were simultaneously detected using stripping chronopotentiometry at the bismuth film electrode. Detection limits of 8 and 10 ppb were obtained for cadmium(II) and lead(II), respectively, for a deposition time of 120 s. The developed method was applied to the determination of lead and cadmium in soils extracts and wastewaters obtained from polluted sites. For comparison purposes, a mercury film electrode and ICP-MS were also used for validation.
引用
收藏
页码:770 / 775
页数:6
相关论文
共 29 条
[1]   Portable fractionation device and screen-printed electrode for measurement of reactive aluminium in environmental samples [J].
Akhtar, P ;
Devereaux, HA ;
Downard, AJ ;
O'Sullivan, B ;
Powell, KJ .
ANALYTICA CHIMICA ACTA, 1999, 381 (01) :49-59
[2]   Immunoassays for metal ions [J].
Blake, DA ;
Blake, RC ;
Khosraviani, M ;
Pavlov, AR .
ANALYTICA CHIMICA ACTA, 1998, 376 (01) :13-19
[3]   Screen-printed anodic stripping voltammetric sensor containing HgO for heavy metal analysis [J].
Choi, JY ;
Seo, K ;
Cho, SR ;
Oh, JR ;
Kahng, SH ;
Park, J .
ANALYTICA CHIMICA ACTA, 2001, 443 (02) :241-247
[4]   Monitoring of heavy metals in urban snow as indicator of atmosphere pollution [J].
Elík, A .
INTERNATIONAL JOURNAL OF ENVIRONMENTAL ANALYTICAL CHEMISTRY, 2002, 82 (01) :37-45
[5]  
Honeychurch KC, 2000, ELECTROANAL, V12, P171, DOI 10.1002/(SICI)1521-4109(200002)12:3<171::AID-ELAN171>3.0.CO
[6]  
2-Q
[7]  
Honeychurch KC, 2002, ELECTROANAL, V14, P177, DOI 10.1002/1521-4109(200202)14:3<177::AID-ELAN177>3.0.CO
[8]  
2-V
[9]   Voltammetric studies of lead at calixarene modified screen-printed carbon electrodes and its trace determination in water by stripping voltammetry [J].
Honeychurch, KC ;
Hart, JP ;
Cowell, DC ;
Arrigan, DWM .
SENSORS AND ACTUATORS B-CHEMICAL, 2001, 77 (03) :642-652
[10]   Voltammetric studies of lead at a 1-(2-pyridylazo)-2-naphthol modified screen-printed carbon electrode and its trace determination in water by stripping voltammetry [J].
Honeychurch, KC ;
Hart, JP ;
Cowell, DC .
ANALYTICA CHIMICA ACTA, 2001, 431 (01) :89-99