A polymeric junction membrane for solid-state reference electrodes

被引:39
作者
Ha, J
Martin, SM
Jeon, Y
Yoon, IJ
Brown, RB
Nam, H
Cha, GS
机构
[1] Kwangwoon Univ, Dept Chem, Chem Sensor Res Grp, Seoul 139701, South Korea
[2] Univ Michigan, Ctr Wireless Integrated MicroSyst, Ann Arbor, MI 48109 USA
[3] Univ Michigan, Ann Arbor, MI 48109 USA
[4] Univ Utah, Sch Elect Engn & Comp Sci, Salt Lake City, UT 84112 USA
基金
美国国家科学基金会;
关键词
ion-selective electrode; reference electrode; liquid junction; cellulose acetate; polyurethane;
D O I
10.1016/j.aca.2005.06.011
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A polymer membrane system composed of 30 wt.% cellulose acetate (CA) and polyurethane (PU) has been developed for the fabrication of mass producible solid-state reference electrodes. The CA/PU membrane exhibits appropriate adhesion to common substrates (e.g., ceramic and silicon chips) of miniaturized electrochemical sensors, allows quick hydration of internal hydrogel layer Q M KCl with a 6 wt.% water soluble polymer) deposited on micro-patterned electrodes resulting in fast preconditioning time (similar to 100 s), and provides a stable reference potential for an extended period (use-lifetime: 25-90 min) by limiting the diffusion of internal electrolytes. The potentiometric responses of the CA/PU membrane-based solid-state reference electrodes drifted rapidly (40-9OmV/h) after their use-lifetime; the results suggest that the internal electrolyte slowly fills the micro-channels of the outer membrane while maintaining stable potential, and begins to diffuse away at an increased rate from the membrane/aqueous solution interface. The potentiometric responses of the polymer membrane-based ion-selective electrodes (for Na+, K+ , Ca2+, Cl- and H+) formed on the same chip with the CA/PU membrane-coated Ag/AgCl reference electrode were examined both in aqueous and physiological samples; their analytical performance closely matched that of sensors measured against a conventional reference electrode. (c) 2005 Elsevier B.V. All rights reserved.
引用
收藏
页码:59 / 66
页数:8
相关论文
共 34 条
[1]   Photo-cross-linked decyl methacrylate films for electrochemical and optical polyion probes [J].
Ambrose, TM ;
Meyerhoff, ME .
ANALYTICAL CHEMISTRY, 1997, 69 (20) :4092-4098
[2]  
Bakker E, 1999, ELECTROANAL, V11, P788, DOI 10.1002/(SICI)1521-4109(199907)11:10/11<788::AID-ELAN788>3.3.CO
[3]  
2-W
[4]  
BARD AJ, 2000, FUNDAMENTALS APPL, pCH2
[5]  
COLLINS SD, 1993, SENSOR ACTUAT B-CHEM, V10, P178
[6]   FIELD-EFFECT TRANSISTOR AS A SOLID-STATE REFERENCE ELECTRODE [J].
COMTE, PA ;
JANATA, J .
ANALYTICA CHIMICA ACTA, 1978, 101 (02) :247-252
[7]   Thick film silver silver chloride reference electrodes [J].
Cranny, AWJ ;
Atkinson, JK .
MEASUREMENT SCIENCE AND TECHNOLOGY, 1998, 9 (09) :1557-1565
[8]   Evaluation of miniaturised solid state reference electrodes on a silicon based component [J].
Desmond, D ;
Lane, B ;
Alderman, J ;
Glennon, JD ;
Diamond, D ;
Arrigan, DWM .
SENSORS AND ACTUATORS B-CHEMICAL, 1997, 44 (1-3) :389-396
[9]   EVALUATION OF A NEW SOLID-STATE REFERENCE ELECTRODE JUNCTION MATERIAL FOR ION-SELECTIVE ELECTRODES [J].
DIAMOND, D ;
MCENROE, E ;
MCCARRICK, M ;
LEWENSTAM, A .
ELECTROANALYSIS, 1994, 6 (11-12) :962-971
[10]   Towards a solid state reference electrode [J].
Eine, K ;
Kjelstrup, S ;
Nagy, K ;
Syverud, K .
SENSORS AND ACTUATORS B-CHEMICAL, 1997, 44 (1-3) :381-388