Comparison of a multiplexed fluorescent covalent microsphere immunoassay and an enzyme-linked immunosorbent assay for measurement of human immunoglobulin G antibodies to anthrax toxins

被引:73
作者
Biagini, RE
Sammons, DL
Smith, JP
MacKenzie, BA
Striley, CAF
Semenova, V
Steward-Clark, E
Stamey, K
Freeman, AE
Quinn, CP
Snawder, JE
机构
[1] NIOSH, CDC,Div Appl Res & Technol, Biol Monitoring Lab Sect,Robert A Taft Labs, Biomonitoring & Hlth Assement Branch, Cincinnati, OH 45226 USA
[2] Ctr Dis Control & Prevent, Natl Ctr Infect Dis, Microbial Pathogenesis & Immune Response Lab, Atlanta, GA USA
关键词
D O I
10.1128/CDLI.11.1.50-55.2004
中图分类号
R392 [医学免疫学]; Q939.91 [免疫学];
学科分类号
100102 ;
摘要
Recently, the Centers for Disease Control and Prevention reported an accurate, sensitive, specific, reproducible, and quantitative enzyme-linked immunosorbent assay (ELISA) for immunoglobulin G (IgG) antibodies to Bacillus anthracis protective antigen (PA) in human serum (C. P. Quinn, V. A. Semenova, C. M. Elie et al., Emerg. Infect. Dis. 8:1103-1110, 2002). The ELISA had a minimum detectable concentration (MDC) of 0.06 mug/ml, which, when dilution adjusted, yielded a whole-serum MDC of 3.0 mug of anti-PA IgG per ml. The reliable detection limit (RDL) was 0.09 mug/ml, while the dynamic range was 0.06 to 1.7 mug/ml. The diagnostic sensitivity of the assay was 97.6% and the diagnostic specificity was 94.2% for clinically verified cases of anthrax. A competitive inhibition anti-PA IgG ELISA was also developed to enhance the diagnostic specificity to 100%. We report a newly developed fluorescence covalent microbead immunosorbent assay (FCMIA) for B. anthracis PA which was Luminex xMap technology. The FCMIA MDC was 0.006 mug of anti-PA IgG per ml, the RDL was 0.016 mug/ml, and the whole-serum equivalent MDC was 1.5 mug/ml. The dynamic range was 0.006 to 6.8 mug/ml. Using this system, we analyzed 20 serum samples for anti-PA IgG and compared our results to those measured by ELISA in a double-masked analysis. The two methods had a high positive correlation (r(2) = 0.852; P < 0.001). The FCMIA appears to have benefits over the ELISA for the measurement of anti-PA IgG, including greater sensitivity and speed, enhanced dynamic range and reagent stability, the use of smaller sample volumes, and the ability to be multiplexed (measurement of more than one analyte simultaneously), as evidenced by the multiplexed measurement in the present report of anti-PA and anti-lethal factor IgG in serum from a confirmed clinical anthrax infection.
引用
收藏
页码:50 / 55
页数:6
相关论文
共 19 条
  • [1] THE USE OF NONLINEAR REGRESSION-ANALYSIS AND THE F-TEST FOR MODEL DISCRIMINATION WITH DOSE-RESPONSE CURVES AND LIGAND-BINDING DATA
    BARDSLEY, WG
    MCGINLAY, PB
    [J]. JOURNAL OF THEORETICAL BIOLOGY, 1987, 126 (02) : 183 - 201
  • [2] Method for simultaneous measurement of antibodies to 23 pneumococcal capsular polysaccharides
    Biagini, RE
    Schlottmann, SA
    Sammons, DL
    Smith, JP
    Snawder, JC
    Striley, CAF
    MacKenzie, BA
    Weissman, DN
    [J]. CLINICAL AND DIAGNOSTIC LABORATORY IMMUNOLOGY, 2003, 10 (05) : 744 - 750
  • [3] Development of multiplexed fluorescence microbead covalent assays (FMCAs) for pesticide biomonitoring
    Biagini, RE
    Murphy, DM
    Sammons, DL
    Smith, JP
    Striley, CAF
    MacKenzie, BA
    [J]. BULLETIN OF ENVIRONMENTAL CONTAMINATION AND TOXICOLOGY, 2002, 68 (04) : 470 - 477
  • [4] Simultaneous detection of 15 human cytokines in a single sample of stimulated peripheral blood mononuclear cells
    de Jager, W
    te Velthuis, H
    Prakken, BJ
    Kuis, W
    Rijkers, GT
    [J]. CLINICAL AND DIAGNOSTIC LABORATORY IMMUNOLOGY, 2003, 10 (01) : 133 - 139
  • [5] Hornung R.W., 1990, Appl. Occupat. Environ. Hygiene, V5, P46, DOI DOI 10.1080/1047322X.1990.10389587
  • [6] ENZYME-LINKED-IMMUNOSORBENT-ASSAY USING A RECOMBINANT BACULOVIRUS-EXPRESSED BACILLUS-ANTHRACIS PROTECTIVE ANTIGEN (PA) - MEASUREMENT OF HUMAN ANTI-PA ANTIBODIES
    IACONOCONNORS, LC
    NOVAK, J
    ROSSI, C
    MANGIAFICO, J
    KSIAZEK, T
    [J]. CLINICAL AND DIAGNOSTIC LABORATORY IMMUNOLOGY, 1994, 1 (01) : 78 - 82
  • [7] Validation of serological assays for diagnosis of infectious diseases
    Jacobson, RH
    [J]. REVUE SCIENTIFIQUE ET TECHNIQUE-OFFICE INTERNATIONAL DES EPIZOOTIES, 1998, 17 (02): : 469 - 486
  • [9] LEPPLA SH, 1991, METHOD ENZYMOL, V195, P153
  • [10] Nix B, 2001, IMMUNOASSAY HDB, P198