A quantitative appraisal of interference by icodextrin metabolites in point-of-care glucose analyses

被引:13
作者
Apperloo, JJ [1 ]
Vader, HL [1 ]
机构
[1] Maxima Med Ctr, Clin Chem Lab, NL-5500 MB Veldhoven, Netherlands
关键词
glucose specificity; icodextrin metabolites; peritoneal dialysis; point-of-care glucose analyzer;
D O I
10.1515/CCLM.2005.054
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
An important number of patients dependent on renal dialysis prefer peritoneal dialysis to hemodialysis. In the case of peritoneal dialysis the glucose polymer icodextrin is frequently added to the dialysis fluid as an osmotic agent, since this polymer is able to maintain an osmotic gradient across the peritoneal membrane longer than monomeric glucose, leading to a prolonged effective ultrafiltration time. It was previously shown that icodextrin is partly able to enter the blood via the lymphatic system, where hydrolysis to glucose oligomers such as maltose and maltotriose occurs. The presence of these oligomers in the blood appears to cause significant overestimations of the glucose values in several point-of-care (POC) glucose analyzers, with potentially dramatic consequences. This effect has been investigated for a series of POC glucose analyzers, both by analyzing the blood of peritoneal dialysis patients and by an in vitro investigation of the quantitative effects of maltose and maltotriose. In particular, POC analyzers utilizing the bacterially produced enzyme glucose dehydrogenase seem to lack glucose specificity.
引用
收藏
页码:314 / 318
页数:5
相关论文
共 9 条
[1]  
DAVIS ALV, 1994, AFR ENTOMOL, V2, P45
[2]  
de Waart DR, 2001, PERITON DIALYSIS INT, V21, P269
[3]  
Janssen W, 1998, CLIN CHEM, V44, P2379
[4]   Global burden of diabetes, 1995-2025 - Prevalence, numerical estimates, and projections [J].
King, H ;
Aubert, RE ;
Herman, WH .
DIABETES CARE, 1998, 21 (09) :1414-1431
[5]   Important causes of hypoglycaemia in patients with diabetes on peritoneal dialysis [J].
Mehmet, S ;
Quan, G ;
Thomas, S ;
Goldsmith, D .
DIABETIC MEDICINE, 2001, 18 (08) :679-682
[6]   Blood glucose overestimation in diabetic patients on continuous ambulatory peritoneal dialysis for end-stage renal disease [J].
Oyibo, SO ;
Pritchard, GM ;
Mclay, L ;
James, E ;
Laing, I ;
Gokal, R ;
Boulton, AJM .
DIABETIC MEDICINE, 2002, 19 (08) :693-696
[7]   Lesson of the week - Spurious hyperglycaemia and icodextrin in peritoneal dialysis fluid [J].
Riley, SG ;
Chess, J ;
Donovan, KL ;
Williams, JD .
BMJ-BRITISH MEDICAL JOURNAL, 2003, 327 (7415) :608-609
[8]  
Wang R, 2004, PERITON DIALYSIS INT, V24, P296
[9]  
Wens R, 1998, PERITON DIALYSIS INT, V18, P603