Use of electrochemical sensors for measurement of air pollution: correcting interference response and validating measurements

被引:196
作者
Cross, Eben S. [1 ]
Williams, Leah R. [1 ]
Lewis, David K. [1 ,2 ]
Magoon, Gregory R. [1 ]
Onasch, Timothy B. [1 ]
Kaminsky, Michael L. [3 ]
Worsnop, Douglas R. [1 ]
Jayne, John T. [1 ]
机构
[1] Aerodyne Res Inc, Ctr Aerosol & Cloud Chem, Billerica, MA 01821 USA
[2] Connecticut Coll, Dept Chem, New London, CT 06320 USA
[3] MIT, 77 Massachusetts Ave, Cambridge, MA 02139 USA
关键词
LOW-COST; PARTICLE MONITORS; QUALITY; PERFORMANCE;
D O I
10.5194/amt-10-3575-2017
中图分类号
P4 [大气科学(气象学)];
学科分类号
0706 ; 070601 ;
摘要
The environments in which we live, work, and play are subject to enormous variability in air pollutant concentrations. To adequately characterize air quality (AQ), measurements must be fast (real time), scalable, and reliable (with known accuracy, precision, and stability over time). Lower-cost air-quality-sensor technologies offer new opportunities for fast and distributed measurements, but a persistent characterization gap remains when it comes to evaluating sensor performance under realistic environmental sampling conditions. This limits our ability to inform the public about pollution sources and inspire policy makers to address environmental justice issues related to air quality. In this paper, initial results obtained with a recently developed lower-cost air-quality-sensor system are reported. In this project, data were acquired with the ARISense integrated sensor package over a 4.5-month time interval during which the sensor system was co-located with a state-operated (Massachusetts, USA) air quality monitoring station equipped with reference instrumentation measuring the same pollutant species. This paper focuses on validating electrochemical (EC) sensor measurements of CO, NO, NO2, and O-3 at an urban neighborhood site with pollutant concentration ranges (parts per billion by volume, ppb; 5 min averages, +/- 1 sigma) : [CO] = 231 +/- 116 ppb (spanning 84-1706 ppb), [NO] = 6.1 +/- 11.5 ppb (spanning 0-209 ppb), [NO2] = 11.7 +/- 8.3 ppb (spanning 0-71 ppb), and [O-3] = 23.2 +/- 12.5 ppb (spanning 0-99 ppb). Through the use of high-dimensional model representation (HDMR), we show that interference effects derived from the variable ambient gas concentration mix and changing environmen-tal conditions over three seasons (sensor flow-cell temperature = 23.4 +/- 8.5 degrees C, spanning 4.1 to 45.2 degrees C; and relative humidity = 50.1 +/- 15.3 %, spanning 9.8-79.9 %) can be effectively modeled for the Alphasense CO-B4, NO-B4, NO2-B43F, and Ox-B421 sensors, yielding (5 min average) root mean square errors (RMSE) of 39.2, 4.52, 4.56, and 9.71 ppb, respectively. Our results substantiate the potential for distributed air pollution measurements that could be enabled with these sensors.
引用
收藏
页码:3575 / 3588
页数:14
相关论文
共 26 条
[1]  
[Anonymous], 2014, BURD DIS AMB HOUS AI
[2]  
Backus A., 2012, HLTH ENV COMMUNITY T
[3]   Assessment of air quality microsensors versus reference methods: The EuNetAir joint exercise [J].
Borrego, C. ;
Costa, A. M. ;
Ginja, J. ;
Amorim, M. ;
Coutinho, M. ;
Karatzas, K. ;
Sioumis, Th. ;
Katsifarakis, N. ;
Konstantinidis, K. ;
De Vito, S. ;
Esposito, E. ;
Smith, P. ;
Andre, N. ;
Gerard, P. ;
Francis, L. A. ;
Castell, N. ;
Schneider, P. ;
Viana, M. ;
Minguillon, M. C. ;
Reimringer, W. ;
Otjes, R. P. ;
von Sicard, O. ;
Pohle, R. ;
Elen, B. ;
Suriano, D. ;
Pfister, V. ;
Prato, M. ;
Dipinto, S. ;
Penza, M. .
ATMOSPHERIC ENVIRONMENT, 2016, 147 :246-263
[4]   Can commercial low-cost sensor platforms contribute to air quality monitoring and exposure estimates? [J].
Castell, Nuria ;
Dauge, Franck R. ;
Schneider, Philipp ;
Vogt, Matthias ;
Lerner, Uri ;
Fishbain, Barak ;
Broday, David ;
Bartonova, Alena .
ENVIRONMENT INTERNATIONAL, 2017, 99 :293-302
[5]  
EPA, 2017, EV EM AIR POL SENS P
[6]  
Hagan D. H., 2017, ATMOS MEAS IN PRESS
[7]   Feasibility of using low-cost portable particle monitors for measurement of fine and coarse particulate matter in urban ambient air [J].
Hana, Inkyu ;
Symanski, Elaine ;
Stock, Thomas H. .
JOURNAL OF THE AIR & WASTE MANAGEMENT ASSOCIATION, 2017, 67 (03) :330-340
[8]   Field calibrations of a low-cost aerosol sensor at a regulatory monitoring site in California [J].
Holstius, D. M. ;
Pillarisetti, A. ;
Smith, K. R. ;
Seto, E. .
ATMOSPHERIC MEASUREMENT TECHNIQUES, 2014, 7 (04) :1121-1131
[9]   Community Air Sensor Network (CAIRSENSE) project: evaluation of low-cost sensor performance in a suburban environment in the southeastern United States [J].
Jiao, Wan ;
Hagler, Gayle ;
Williams, Ronald ;
Sharpe, Robert ;
Brown, Ryan ;
Garver, Daniel ;
Judge, Robert ;
Caudill, Motria ;
Rickard, Joshua ;
Davis, Michael ;
Weinstock, Lewis ;
Zimmer-Dauphinee, Susan ;
Buckley, Ken .
ATMOSPHERIC MEASUREMENT TECHNIQUES, 2016, 9 (11) :5281-5292
[10]   The rise of low-cost sensing for managing air pollution in cities [J].
Kumar, Prashant ;
Morawska, Lidia ;
Martani, Claudio ;
Biskos, George ;
Neophytou, Marina ;
Di Sabatino, Silvana ;
Bell, Margaret ;
Norford, Leslie ;
Britter, Rex .
ENVIRONMENT INTERNATIONAL, 2015, 75 :199-205