A biosensor array based on polyaniline

被引:195
作者
Sangodkar, H
Sukeerthi, S
Srinivasa, RS
Lal, R
Contractor, AQ
机构
[1] INDIAN INST TECHNOL,DEPT CHEM,BOMBAY 400076,MAHARASHTRA,INDIA
[2] INDIAN INST TECHNOL,DEPT MET ENGN & MAT SCI,BOMBAY 400076,MAHARASHTRA,INDIA
[3] INDIAN INST TECHNOL,DEPT ELECT ENGN,BOMBAY 400076,MAHARASHTRA,INDIA
关键词
D O I
10.1021/ac950655w
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
This paper describes the fabrication of polyaniline-based microsensors and microsensor arrays for the estimation of glucose, urea, and triglycerides. Microelectronics technology has been used to produce gold interdigitated microelectrodes on oxidized silicon wafers, Polymer deposition and enzyme immobilization has been done electrochemically. Electrochemical potential control has been used to direct enzyme immobilization to the chosen microelectrodes and prevent it at other microelectrodes in contact with the enzyme solution. This has enabled the immobilization of three different enzymes on three closely spaced microelectrodes, resulting in a sensor array which can analyze a sample containing a mixture of glucose, urea, and triolein in a single measurement using a few microliters of the sample, This strategy is quite general and can be extended to other enzyme-substrate systems to eventually produce an ''electronic tongue''.
引用
收藏
页码:779 / 783
页数:5
相关论文
共 12 条
[1]   A MICROELECTROCHEMICAL ENZYME TRANSISTOR RESPONSIVE TO GLUCOSE [J].
BARTLETT, PN ;
BIRKIN, PR .
ANALYTICAL CHEMISTRY, 1994, 66 (09) :1552-1559
[2]   CONDUCTING POLYMER-BASED BIOSENSORS [J].
CONTRACTOR, AQ ;
SURESHKUMAR, TN ;
NARAYANAN, R ;
SUKEERTHI, S ;
LAL, R ;
SRINIVASA, RS .
ELECTROCHIMICA ACTA, 1994, 39 (8-9) :1321-1324
[3]   ENZYME ENTRAPMENT IN ELECTRICALLY CONDUCTING POLYMERS - IMMOBILIZATION OF GLUCOSE-OXIDASE IN POLYPYRROLE AND ITS APPLICATION IN AMPEROMETRIC GLUCOSE SENSORS [J].
FOULDS, NC ;
LOWE, CR .
JOURNAL OF THE CHEMICAL SOCIETY-FARADAY TRANSACTIONS I, 1986, 82 :1259-1264
[4]  
GHOLAMIAN M, 1986, P INDIAN AS-CHEM SCI, V97, P457
[5]   BIOSENSOR BASED ON CONDUCTING POLYMERS [J].
HOA, DT ;
KUMAR, TNS ;
PUNEKAR, NS ;
SRINIVASA, RS ;
LAL, R ;
CONTRACTOR, AQ .
ANALYTICAL CHEMISTRY, 1992, 64 (21) :2645-2646
[6]   SIMULTANEOUS IMMOBILIZATION OF GLUCOSE-OXIDASE AND A MEDIATOR IN CONDUCTING POLYMER-FILMS [J].
IWAKURA, C ;
KAJIYA, Y ;
YONEYAMA, H .
JOURNAL OF THE CHEMICAL SOCIETY-CHEMICAL COMMUNICATIONS, 1988, (15) :1019-1020
[7]   CHEMICAL DERIVATIZATION OF MICROELECTRODE ARRAYS BY OXIDATION OF PYRROLE AND N-METHYLPYRROLE - FABRICATION OF MOLECULE-BASED ELECTRONIC DEVICES [J].
KITTLESEN, GP ;
WHITE, HS ;
WRIGHTON, MS .
JOURNAL OF THE AMERICAN CHEMICAL SOCIETY, 1984, 106 (24) :7389-7396
[8]   POLYANILINE - PROTONIC ACID DOPING TO THE METALLIC REGIME [J].
MACDIARMID, AG ;
CHIANG, JC ;
HUANG, WS ;
HUMPHREY, BD ;
SOMASIRI, NLD .
MOLECULAR CRYSTALS AND LIQUID CRYSTALS, 1985, 125 (1-4) :309-318
[9]   2-DIMENSIONAL IMAGING OF O-2, H2O2, AND GLUCOSE DISTRIBUTIONS BY AN ARRAY OF 400 INDIVIDUALLY ADDRESSABLE MICROELECTRODES [J].
MEYER, H ;
DREWER, H ;
GRUNDIG, B ;
CAMMANN, K ;
KAKEROW, R ;
MANOLI, Y ;
MOKWA, W ;
ROSPERT, M .
ANALYTICAL CHEMISTRY, 1995, 67 (07) :1164-1170
[10]   PENICILLIN SENSOR BASED ON A MICROARRAY ELECTRODE COATED WITH PH-RESPONSIVE POLYPYRROLE [J].
NISHIZAWA, M ;
MATSUE, T ;
UCHIDA, I .
ANALYTICAL CHEMISTRY, 1992, 64 (21) :2642-2644