Comparison of different 3D navigation systems by a clinical "user"

被引:24
作者
Cartellieri, M
Kremser, J
Vorbeck, F
机构
[1] Univ Vienna, Gen Hosp, Dept Otorhinolaryngol, A-1090 Vienna, Austria
[2] Univ Vienna, Dept Radiol, A-1090 Vienna, Austria
关键词
3D navigation system; system accuracy;
D O I
10.1007/s004050000302
中图分类号
R76 [耳鼻咽喉科学];
学科分类号
100213 ;
摘要
Three-dimensional navigation systems are routinely used in endoscopic skull base surgery neurosurgery, maxillo-facial and endoscopic sinus surgery. Their precision can, however, change in the course of one experiment. We have compared five different 3D navigation systems and discuss here possible reasons for the limits of system precision. A plexiglass cube an which test points were marked served as a test-model. Two well-trained system users measured the distances between the test points in each of the five systems. The results were compared with reference data provided by the NUMEREX(TM) device at the Technical University of Vienna. The accuracy data shown by all these 3D navigation systems ranged from 0.0 mm to 6.67 mm. The accuracy data of a system calculated in advance did not always correspond with the system precision on the screen. The system precision in the center of the cube was higher than on its surface, which made us conclude that the angle between the tracker system and the pointing device touching the test point may be critical for system precision. Applying an automatic registration step did not result in greater system precision. Slice thickness and the angle of the pointing device seem to be responsible for system precision.
引用
收藏
页码:38 / 41
页数:4
相关论文
共 16 条
  • [1] Computer-aided endoscopic sinus surgery
    Anon, JB
    [J]. LARYNGOSCOPE, 1998, 108 (07) : 949 - 961
  • [2] BRENT A, 1997, ARCH OTOLARYNGOL, V123, P706
  • [3] COMPUTER-ASSISTED FRONTAL SINUSOTOMY
    CARRAU, RL
    SNYDERMAN, CH
    CURTIN, HB
    WEISSMAN, JL
    [J]. OTOLARYNGOLOGY-HEAD AND NECK SURGERY, 1994, 111 (06) : 727 - 732
  • [4] Image-guided endoscopic surgery: Results of accuracy and performance in a multicenter clinical study using an electromagnetic tracking system
    Fried, MP
    Kleefield, J
    Gopal, H
    Reardon, E
    Ho, BT
    Kuhn, FA
    [J]. LARYNGOSCOPE, 1997, 107 (05) : 594 - 601
  • [5] FRIED MP, 1997, CURR OPIN OTOLARYNGO, V5, P22
  • [6] GALLOWAY R L JR, 1991, Biomedical Instrumentation and Technology, V25, P457
  • [7] GOLDSMITH MM, 1995, AM J OTOL, V16, P475
  • [8] Computer-assisted surgery in the frontal and maxillary sinus
    Gunkel, AR
    Freysinger, W
    Thumfart, WF
    [J]. LARYNGOSCOPE, 1997, 107 (05) : 631 - 633
  • [9] AN ACCURATE AND ERGONOMIC METHOD OF REGISTRATION FOR IMAGE-GUIDED NEUROSURGERY
    HENDERSON, JM
    SMITH, KR
    BUCHOLZ, RD
    [J]. COMPUTERIZED MEDICAL IMAGING AND GRAPHICS, 1994, 18 (04) : 273 - 277
  • [10] HUDGINS PA, 1993, RADIOL CLIN N AM, V31, P21