Variational formulation of open-ended coaxial line in contact with layered biological medium

被引:83
作者
Alanen, E [1 ]
Lahtinen, T
Nuutinen, J
机构
[1] Kuopio Univ Hosp, Dept Oncol, FIN-70210 Kuopio, Finland
[2] Univ Kuopio, Dept Appl Phys, FIN-70210 Kuopio, Finland
关键词
coaxial probe; skin impedance; variational method;
D O I
10.1109/10.720202
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
An open-ended coaxial probe designed to measure layered biological media is analyzed with a new method. The probe is considered as an electrostatic circuit element whose capacitance is solved using a stationary functional. The fundamental transverse electric and magnetic field (TEM)-mode and the series of evanescent wavemodes in the coaxial cable are used as basis functions. The field-outside the probe is solved using a Hankel transform, The capacitance is calculated for homogeneous materials and two-layer structures and the results are compared with values measured with a phantom model. The method can be easily extended for structures with an arbitrary number of layers. A practical approximation for two-layer cases, originally developed to take into account the effect of subcutaneous fat in skin measurements, is presented and its validity for different combinations of dielectric constants and the thickness of the first layer is demonstrated, The static approximation limits the frequency range, but it covers biological measurements up to 500 MHz, The developed method is accurate and easy to adopt in practice.
引用
收藏
页码:1241 / 1248
页数:8
相关论文
共 26 条
[1]  
[Anonymous], 1980, Gos. Izd-vo Fiz.-Mat. Literatury
[2]  
Arfken G, 1985, MATH METHODS PHYS, P573
[3]   MEASUREMENT OF RADIO-FREQUENCY PERMITTIVITY OF BIOLOGICAL TISSUES WITH AN OPEN-ENDED COAXIAL LINE .1. [J].
ATHEY, TW ;
STUCHLY, MA ;
STUCHLY, SS .
IEEE TRANSACTIONS ON MICROWAVE THEORY AND TECHNIQUES, 1982, 30 (01) :82-86
[4]   ANALYSIS OF RADIATION FROM AN OPEN-ENDED COAXIAL LINE INTO STRATIFIED DIELECTRICS [J].
BAKHTIARI, S ;
GANCHEV, SI ;
ZOUGHI, R .
IEEE TRANSACTIONS ON MICROWAVE THEORY AND TECHNIQUES, 1994, 42 (07) :1261-1267
[5]   MEASUREMENT OF LOW-PERMITTIVITY MATERIALS BASED ON A SPECTRAL-DOMAIN ANALYSIS FOR THE OPEN-ENDED COAXIAL PROBE [J].
DELANGHE, P ;
BLOMME, K ;
MARTENS, L ;
DEZUTTER, D .
IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 1993, 42 (05) :879-886
[6]   ANALYSIS OF THE OPEN-ENDED COAXIAL LINE SENSORS [J].
DELECKI, ZA ;
STUCHLY, SS .
JOURNAL OF ELECTROMAGNETIC WAVES AND APPLICATIONS, 1990, 4 (02) :169-183
[7]   STATIC ANALYSIS OF AN OPEN-ENDED COAXIAL LINE TERMINATED BY LAYERED MEDIA [J].
FAN, S ;
STAEBELL, K ;
MISRA, D .
IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 1990, 39 (02) :435-437
[8]  
Foster K.R., 1986, CRC HDB BIOL EFFECTS, P27
[9]   The dielectric properties of biological tissues .3. Parametric models for the dielectric spectrum of tissues [J].
Gabriel, S ;
Lau, RW ;
Gabriel, C .
PHYSICS IN MEDICINE AND BIOLOGY, 1996, 41 (11) :2271-2293
[10]   NUMERICAL-ANALYSIS OF OPEN-ENDED CO-AXIAL LINES [J].
GAJDA, GB ;
STUCHLY, SS .
IEEE TRANSACTIONS ON MICROWAVE THEORY AND TECHNIQUES, 1983, 31 (05) :380-384