Recalibration of the point-of-care test for CARDIAC T Quantitative with Elecsys Troponin T 3rd generation

被引:18
作者
Collinson, PO
Jorgensen, B
Sylvén, C
Haass, M
Chwallek, F
Katus, HA
Müller-Bardorff, M
Derhaschnig, U
Hirschl, MM
Zerback, R
机构
[1] Roche Diagnost GmbH, Evaluat Dept Patient Care, D-68305 Mannheim, Germany
[2] Mayday Univ Hosp, Dept Pathol, London, England
[3] Alborg Hosp, Dept Clin Biochem, Alborg, Denmark
[4] Huddinge Hosp, Dept Cardiol, S-14186 Huddinge, Sweden
[5] Univ Heidelberg, Med Klin, D-6900 Heidelberg, Germany
[6] Med Univ Lubeck, Med Klin 2, D-23538 Lubeck, Germany
[7] Allgemeines Krankenhaus Univ Kliniken, Vienna, Austria
关键词
cardiac troponin T; point-of-care testing; calibration; acute myocardial infarction; minor myocardial injury; multicentre evaluation;
D O I
10.1016/S0009-8981(01)00428-4
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
The rapid troponin T assay CARDIAC T Quantitative was recalibrated using Elecsys Troponin T 3rd Generation as a new reference method. This paper presents the method comparisons at six centres using the new reference method. Method comparison between CARDIAC T Quantitative Versus Elecsys Troponin T 3rd Generation were performed using 319 samples from patients with acute coronary syndromes. The quality of the CARDIAC T Quantitative was controlled by a daily single determination of CARDIAC Control Troponin T, and for the Elecsys Troponin T 3rd Generation, the Elecsys controls were included in each run. The results for the control materials for the CARDIAC T Quantitative were between 93% and 107% of the target Values. The CV ranged from 7% to 16%. From the regression analysis, according to Bablok and Passing (y = 1.07x) and the Bland and Altman plot, the bias between CARDIAC T Quantitative and Elecsys Troponin T 3rd Generation is from +6% to +7%, The correlation coefficient is 0.93, and a 3 x 3 comparison of the clinical efficiency yielded 92% clinical concordance between CARDIAC T Quantitative and Elecsys Troponin T 3rd Generation. In conclusion, CARDIAC T Quantitative was in good agreement with the reference and calibration method Elecsys Troponin T 3rd Generation. (C) 2001 Published by Elsevier Science B.V.
引用
收藏
页码:197 / 203
页数:7
相关论文
共 6 条
  • [1] BRIMHALL BB, 1995, CLIN CHEM, V41, pS167
  • [2] Collinson PO, 1996, EUR J CLIN CHEM CLIN, V34, P591
  • [3] Relation between troponin T and the risk of subsequent cardiac events in unstable coronary artery disease
    Lindahl, B
    Venge, P
    Wallentin, L
    [J]. CIRCULATION, 1996, 93 (09) : 1651 - 1657
  • [4] Evaluation of a point-of-care system for quantitative determination of troponin T and myoglobin
    Müller-Bardorff, M
    Sylvén, C
    Rasmanis, G
    Jorgensen, B
    Collinson, PO
    Waldenhofer, U
    Hirschl, MM
    Laggner, AN
    Gerhardt, W
    Hafner, G
    Labaere, I
    Leinberger, R
    Zerback, R
    Katus, HA
    [J]. CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2000, 38 (06) : 567 - 574
  • [5] DEVELOPMENT AND CHARACTERIZATION OF A RAPID ASSAY FOR BEDSIDE DETERMINATIONS OF CARDIAC TROPONIN-T
    MULLERBARDORFF, M
    FREITAG, H
    SCHEFFOLD, T
    REMPPIS, A
    KUBLER, W
    KATUS, HA
    [J]. CIRCULATION, 1995, 92 (10) : 2869 - 2875
  • [6] Cardiac troponin T levels for risk stratification in acute myocardial ischemia
    Ohman, EM
    Armstrong, PW
    Christenson, RH
    Granger, CB
    Katus, HA
    Hamm, CW
    OHanesian, MA
    Wagner, GS
    Kleiman, NS
    Harrell, FE
    Califf, RM
    Topol, EJ
    [J]. NEW ENGLAND JOURNAL OF MEDICINE, 1996, 335 (18) : 1333 - 1341