Diffusing fibre tip for the minimally invasive treatment of liver tumours by interstitial laser coagulation (ILC): An experimental ex vivo study

被引:27
作者
Germer, CT
Albrecht, D
Isbert, C
Ritz, J
Roggan, A
Buhr, HJ
机构
[1] Free Univ Berlin, Dept Surg, D-1000 Berlin, Germany
[2] Free Univ Berlin, Dept Laser Med, D-1000 Berlin, Germany
关键词
interstitial laser coagulation; laser applicator; laser induced thermotherapy; liver tumours; Nd-YAG laser;
D O I
10.1007/s101030050018
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
A newly developed diffusing laser applicator was examined for interstitial laser coagulation (ILC) of liver tumours. The applicator consisted of a matted quartz core and a quartz glass dome, also matted on its inner surface and sealed to the fibre. The applicator provided a homogeneous light intensity distribution over an active length of about 20 mm. Lesions were created in an ex-vivo porcine liver model using a Nd-YAG laser comparing the new diffusing tip with a Ringmode(R)-ITT applicator in order to find optimal laser parameters and damage thresholds. The lesions were investigated using macroscopic size measurement, volume calculation and histological examination (H&E, NADPH-dehydrogenase). The damage threshold of the diffusing tip was 6 W at 14 min exposure time whereas the Ringmode(R)-ITT applicator had its limit at 5 W and 12 min exposure. Comparing various exposure times showed that treatment over a time of more than 840 s did not significantly increase the lesion volume. At 5 W and 720 s the mean lesion volume was 6.9 +/- 1.1 cm(3) with the diffusing tip and 6.3 +/- 0.6 cm(3) with the Ringmode(R)-ITT applicator, both having a slight ellipsoidal shape. Hence,the created lesions were not significantly different for both applicators when the same laser parameters were applied. On the other hand, the new diffusing tip had a higher damage threshold and was therefore capable of producing maximal coagulation volumes of up to 7.9 +/- 0.5 cm(3) at 5 W and 20 min. The experiments showed that lesions with a dimension of 31 x 22 mm can be achieved with. the diffusing applicator which seem suitably sized for treating small human liver metastases in a single laser session.
引用
收藏
页码:32 / 39
页数:8
相关论文
共 32 条
[1]   LOCAL TREATMENT OF COLORECTAL LIVER METASTASES - A COMPARISON OF INTERSTITIAL LASER PHOTOCOAGULATION (ILP) AND PERCUTANEOUS ALCOHOL INJECTION (PAI) [J].
AMIN, Z ;
BOWN, SG ;
LEES, WR .
CLINICAL RADIOLOGY, 1993, 48 (03) :166-171
[2]   CYLINDRICAL IRRADIATOR FIBER TIP FOR PHOTODYNAMIC THERAPY [J].
ARNFIELD, M ;
GONZALEZ, S ;
LEA, P ;
TULIP, J ;
MCPHEE, M .
LASERS IN SURGERY AND MEDICINE, 1986, 6 (02) :150-154
[3]  
BALCHUM OJ, 1985, CLIN CHEST MED, V6, P255
[4]   PHOTOTHERAPY OF TUMORS [J].
BOWN, SG .
WORLD JOURNAL OF SURGERY, 1983, 7 (06) :700-709
[5]  
CAVALIER.R, 1967, CANCER, V20, P1351, DOI 10.1002/1097-0142(196709)20:9<1351::AID-CNCR2820200902>3.0.CO
[6]  
2-#
[7]  
CUSCHIERI A, 1995, SURG ENDOSC-ULTRAS, V9, P483
[8]  
FRANK F, 1990, LASERMEDIZIN, V1, P36
[9]   TUMOR-THERAPY WITH HEMATOPORPHYRIN DERIVATIVE AND LASERS VIA A PERCUTANEOUS FIBEROPTIC TECHNIQUE - PRECLINICAL EXPERIMENTS [J].
GATENBY, RA ;
HAMMOND, ND ;
BROWN, DQ .
RADIOLOGY, 1987, 163 (01) :167-171
[10]   Experimental study of laparoscopic laser-induced thermotherapy for liver tumours [J].
Germer, CT ;
Albrecht, D ;
Roggan, A ;
Isbert, C ;
Buhr, HJ .
BRITISH JOURNAL OF SURGERY, 1997, 84 (03) :317-320