Negative interference in cardiac troponin I immunoassays by circulating troponin autoantibodies

被引:101
作者
Eriksson, S
Halenius, H
Pulkki, K
Hellman, J
Pettersson, K
机构
[1] Univ Turku, Dept Biotechnol, FIN-20520 Turku, Finland
[2] Univ Helsinki, Cent Hosp, Dept Lab Diagnost, Helsinki, Finland
关键词
D O I
10.1373/clinchem.2004.040063
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Background: There are numerous potential sources of interference in immunoassays. Our aim was to identify the blood component that causes negative interference in cardiac troponin I (cTnI) immunoassays based on antibodies against the central part of cTnI. Methods: We isolated an interfering factor (IF) from a sample With low recovery of added cTnI, using several consecutive purification steps: caprylic acid precipitation, ammonium sulfate precipitation, and purification on Cibacron Blue gel and protein G columns. Purified IF was identified by gel electrophoresis and mass spectrometric analysis of protein bands. For the direct detection of human antibodies to cardiac troponin in serum samples, we developed immunoassays using three different anti-human immunoglobulin antibodies and measured troponin antibodies in samples with low and normal cTnI recovery. Results: Treatment with caprylic acid did not precipitate IF, but IF precipitated at 40% ammonium sulfate saturation. IF bound to a Cibacron Blue gel column, from which it was eluted with a linear salt gradient; it also bound to protein G. Gel electrophoresis of purified IF showed two major bands with molecular masses corresponding to the heavy (similar to 50 kDa) and light chains (similar to 25 kDa) of immunoglobulin, and their identities were confirmed by mass spectrometry. The presence of troponin-specific autoantibodies was confirmed in samples with low recoveries of cTnI by three different immunoassays. The median signals were significantly higher in 10 samples with low recovery than in 10 samples with normal recovery of cTnI (P <= 0.007). Conclusions: Circulating autoantibodies to cTnI or other proteins of the troponin complex can be a source of negative interference in cTnI immunoassays. (c) 2005 American Association for Clinical Chemistry.
引用
收藏
页码:839 / 847
页数:9
相关论文
共 43 条
[1]   Myocardial infarction redefined -: A consensus Document of the Joint European Society of Cardiology/American College of Cardiology Committee for the Redefinition of Myocardial Infarction [J].
Alpert, JS ;
Antman, E ;
Apple, F ;
Armstrong, PW ;
Bassand, JP ;
de Luna, AB ;
Beller, G ;
Breithardt, G ;
Chaitman, BR ;
Clemmensen, P ;
Falk, E ;
Fishbein, MC ;
Galvani, M ;
Garson, A ;
Grines, C ;
Hamm, C ;
Jaffe, A ;
Katus, H ;
Kjekshus, J ;
Klein, W ;
Klootwijk, P ;
Lenfant, C ;
Levy, D ;
Levy, RI ;
Luepker, R ;
Marcus, F ;
Näslund, U ;
Ohman, M ;
Pahlm, O ;
Poole-Wilson, P ;
Popp, R ;
Alto, P ;
Pyörälä, K ;
Ravkilde, J ;
Rehnquist, N ;
Roberts, W ;
Roberts, R ;
Roelandt, J ;
Rydén, L ;
Sans, S ;
Simoons, ML ;
Thygesen, K ;
Tunstall-Pedoe, H ;
Underwood, R ;
Uretsky, BF ;
Van de Werf, F ;
Voipio-Pulkki, LM ;
Wagner, G ;
Wallentin, L ;
Wijns, W .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2000, 36 (03) :959-969
[2]  
Apple FS, 1999, CLIN CHEM, V45, P18
[3]  
Bohner J, 1996, CLIN CHEM, V42, P2046
[4]  
Campbell Ailsa M., 1996, P95, DOI 10.1016/B978-012214730-2/50006-6
[5]  
Dasgupta A, 1999, AM J CLIN PATHOL, V112, P753
[6]   Negative interference of bilirubin and hemoglobin in the MEIA troponin I assay but not in the MEIA CK-MB assay [J].
Dasgupta, A ;
Wells, A ;
Biddle, DA .
JOURNAL OF CLINICAL LABORATORY ANALYSIS, 2001, 15 (02) :76-80
[7]   PURIFICATION OF DNA ANTIBODIES USING CIBACRON BLUE F3GA AFFINITY-CHROMATOGRAPHY [J].
EMLEN, W ;
BURDICK, G .
JOURNAL OF IMMUNOLOGICAL METHODS, 1983, 62 (02) :205-215
[8]   Comparison of cardiac troponin I immunoassays variably affected by circulating autoantibodies [J].
Eriksson, S ;
Ilva, T ;
Becker, C ;
Lund, J ;
Porela, P ;
Pulkki, K ;
Voipio-Pulkki, LM ;
Pettersson, K .
CLINICAL CHEMISTRY, 2005, 51 (05) :848-855
[9]   An interfering component in cardiac troponin I immunoassays - Its nature and inhibiting effect on the binding of antibodies against different epitopes [J].
Eriksson, S ;
Junikka, M ;
Pettersson, K .
CLINICAL BIOCHEMISTRY, 2004, 37 (06) :472-480
[10]   Negative interference in cardiac troponin I immunoassays from a frequently occurring serum and plasma component [J].
Eriksson, S ;
Junikka, M ;
Laitinen, P ;
Majamaa-Voltti, K ;
Alfthan, H ;
Pettersson, K .
CLINICAL CHEMISTRY, 2003, 49 (07) :1095-1104