Robust registration method for interventional MRI-guided thermal ablation of prostate cancer

被引:8
作者
Fei, BW [1 ]
Wheaton, A [1 ]
Lee, Z [1 ]
Nagano, K [1 ]
Duerk, JL [1 ]
Wilson, DL [1 ]
机构
[1] Case Western Reserve Univ, Dept Biomed Engn, Cleveland, OH 44106 USA
来源
MEDICAL IMAGING 2001: VISUALIZATION, DISPLAY, AND IMAGE-GUIDED PROCEDURES | 2001年 / 4319卷
关键词
image registration; mutual information; magnetic resonance imaging (MRI); interventional MRI; prostate cancer; minimally invasive treatment; thermal ablation;
D O I
10.1117/12.428092
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
We are investigating methods to register live-time interventional magnetic resonance imaging (iMRI) slice images with a previously obtained, high resolution MRI image volume. The immediate application is for iMRI-guided treatments of prostate cancer. We created and evaluated a slice-to-volume mutual information registration algorithm for MR images with special features to improve robustness. Features included a multi-resolution approach and automatic restarting to avoid local minima. We acquired 3D volume images from a 1.5 T MRI system and simulated iMRI images. To assess the quality of registration, we calculated 3D displacement on a voxel-by-voxel basis over a volume of interest between slice-to-volume registration and volume-to-volume registrations that were previously shown to be quite accurate. More than 500 registration experiments were performed on MR images of volunteers. The slice-to-volume registration algorithm was very robust for transverse slice images covering the prostate. A 100% success rate was achieved with an acceptance criterion of < 1.0 mm displacement error over the prostate. Our automatic slice-to-volume mutual information registration algorithm is robust and probably sufficiently accurate to aid in the application of iMRI-guided thermal ablation of prostate cancer.
引用
收藏
页码:53 / 60
页数:8
相关论文
共 25 条
  • [1] OPTIMIZATION OF PROSTATE CARCINOMA STAGING - COMPARISON OF IMAGING AND CLINICAL METHODS
    BONI, RAH
    BONER, JA
    DEBATIN, JF
    TRINKLER, F
    KNONAGEL, H
    VONHOCHSTETTER, A
    HELFENSTEIN, U
    KRESTIN, GP
    [J]. CLINICAL RADIOLOGY, 1995, 50 (09) : 593 - 600
  • [2] Carr T, 2000, AUSTRALAS J AGEING, V19, P3
  • [3] Cox RW, 1999, MAGNET RESON MED, V42, P1014, DOI 10.1002/(SICI)1522-2594(199912)42:6<1014::AID-MRM4>3.0.CO
  • [4] 2-F
  • [5] FEI BW, UNPUB SLICE VOLUME R
  • [6] FEI BW, 2001, UNPUB IEEE T MED IMA
  • [7] Accurate registration of serial 3D MR brain images and its application to visualizing change in neurodegenerative disorders
    Freeborough, PA
    Woods, RP
    Fox, NC
    [J]. JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1996, 20 (06) : 1012 - 1022
  • [8] GRAY H, 1977, GRAY ANAT CLASSIC CO, P1010
  • [9] Herk M, 1995, INT J RADIAT ONCOL, V33, P1311
  • [10] Hill D L, 1997, Comput Aided Surg, V2, P264, DOI 10.3109/10929089709148116