Detection sensitivity of a commercial lung nodule CAD system in a series of pathologically proven lung cancers

被引:24
作者
Godoy, Myrna C. B. [1 ]
Cooperberg, Peter L. [1 ]
Maizlin, Zeev V. [1 ]
Yuan, Ren [2 ]
McWilliams, Annette [3 ]
Lam, Stephen [3 ]
Mayo, John R. [2 ]
机构
[1] Univ British Columbia, Dept Radiol, St Pauls Hosp, Vancouver, BC V6Z 1Y6, Canada
[2] Univ British Columbia, Vancouver Gen Hosp, Dept Radiol, Vancouver, BC V5Z 1M9, Canada
[3] British Columbia Canc Agcy, Canc Imaging Dept, Vancouver, BC V5Z 4E6, Canada
关键词
computer tomography (CT); lung cancer screening; computer-aided detection (CAD);
D O I
10.1097/RTI.0b013e3181339edb
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 [临床医学]; 100207 [影像医学与核医学]; 1009 [特种医学];
摘要
Purpose: To evaluate the performance of a commercially available computer-aided detection (CAD) system in a series of pathologically proven lung cancers. Materials and Methods: Sixty-nine chest computed tomography (CT) scans obtained in 12 subjects (8 females, 4 males, age 51 to 75 y, mean 63 y) with 15 pathologically proven lung cancers were retrospectively selected from 2156 entry and follow-up CT scans from a lung cancer screening program. CT scans were retrospectively analyzed using a commercially available CAD system for detecting lung nodules. Results: When first detectable proven lung cancer nodules ranged in maximum diameter from 3 to 38 mm (10.4 +/- 9.2 mm) with CAD detection sensitivity stratified by size: 0/2 (0%) <= 3mm, 5/8 (62.5%) 4 to 10mm, 2/3 (66.7%) 11 to 15mm, 0/0 16 to 20 mm, 2/2 (100%) > 20 mm, and overall sensitivity 9/15 (60%). The sensitivity for all CT scans (first detectable and follow-up), stratified by nodule size as above, was, respective y, 0/2, 18/25, 24/28, 6/9, 5/5, and overall 53/69 (76.8%). Excluding nodules < 4mm and pure ground-glass nodules, the sensitivity for all CT scans by size was 18/24 (75%) 4 to 10mm, 21/22 (95.4%) 11 to 15mm, 6/6 (100%) 16 to 20mm, 515 (100%) > 20mm, and overall 50/57 (87.7%). At resection (13) or biopsy (2) nodules were: adenocarcinoma (10), squamous cell carcinoma (3), and small cell carcinoma (2). Conclusions: The CAD system showed good sensitivity for solid and semisolid cancers >= 4 mm (sensitivity 87.7%) and excellent for those >= 11 mm (sensitivity > 95.4%).
引用
收藏
页码:1 / 6
页数:6
相关论文
共 14 条
[1]
*AM CAN SOC, 2007, CANC FACTS FIG 2007
[2]
Lung cancer: Performance of automated lung nodule detection applied to cancers missed in a CT screening program [J].
Armato, SG ;
Li, F ;
Giger, ML ;
MacMahon, H ;
Sone, S ;
Doi, K .
RADIOLOGY, 2002, 225 (03) :685-692
[3]
Pulmonary nodules: Automated detection on CT images with morphologic matching algorithm preliminary results [J].
Bae, KT ;
Kim, JS ;
Na, YH ;
Kim, KG ;
Kim, JH .
RADIOLOGY, 2005, 236 (01) :286-294
[4]
Lung micronodules: Automated method for detection at thin-section CT - Initial experience [J].
Brown, MS ;
Goldin, JG ;
Suh, RD ;
McNitt-Gray, MF ;
Sayre, JW ;
Aberle, DR .
RADIOLOGY, 2003, 226 (01) :256-262
[5]
Early Lung Cancer Action Project: overall design and findings from baseline screening [J].
Henschke, CI ;
McCauley, DI ;
Yankelevitz, DF ;
Naidich, DP ;
McGuinness, G ;
Miettinen, OS ;
Libby, DM ;
Pasmantier, MW ;
Koizumi, J ;
Altorki, NK ;
Smith, JP .
LANCET, 1999, 354 (9173) :99-105
[6]
Peripheral lung cancer: Screening and detection with low-dose spiral CT versus radiography [J].
Kaneko, M ;
Eguchi, K ;
Ohmatsu, H ;
Kakinuma, R ;
Naruke, T ;
Suemasu, K ;
Moriyama, N .
RADIOLOGY, 1996, 201 (03) :798-802
[7]
Lung nodule detection on chest CT: Evaluation of a computer-aided detection (CAD) system [J].
Lee, IJ ;
Gamsu, G ;
Czum, J ;
Wu, N ;
Johnson, R ;
Chakrapani, S .
KOREAN JOURNAL OF RADIOLOGY, 2005, 6 (02) :89-93
[8]
Computer-aided detection of peripheral lung cancers missed at CT: ROC analyses without and with localization [J].
Li, F ;
Arimura, H ;
Suzuki, K ;
Shiraishi, J ;
Li, Q ;
Abe, H ;
Engelmann, R ;
Sone, S ;
MacMahon, H ;
Doi, K .
RADIOLOGY, 2005, 237 (02) :684-690
[9]
Lung cancer screening - A different paradigm [J].
McWilliams, A ;
Mayo, J ;
MacDonald, S ;
LeRiche, JC ;
Palcic, B ;
Szabo, E ;
Lam, S .
AMERICAN JOURNAL OF RESPIRATORY AND CRITICAL CARE MEDICINE, 2003, 168 (10) :1167-1173
[10]
Computer-aided diagnosis as a second reader - Spectrum of findings in CT studies of the chest interpreted as normal [J].
Peldschus, K ;
Herzog, P ;
Wood, SA ;
Cheema, JI ;
Costello, P ;
Schoepf, UJ .
CHEST, 2005, 128 (03) :1517-1523