Trade-offs in miniature quadrupole designs

被引:46
作者
Boumsellek, S [1 ]
Ferran, RJ [1 ]
机构
[1] Ferran Sci Inc, San Diego, CA 92121 USA
关键词
D O I
10.1016/S1044-0305(01)00248-3
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Pressing needs for miniature mass spectrometers became apparent during the last decade in process monitoring and control, space exploration, and environmental screening. Besides the small footprint, common requirements include low cost, low power consumption, field portability, reliability, autonomy, and ease-of-use. Design concepts and construction technologies of miniaturized quadrupole sensors were guided by cost reduction requirements without sacrifice of performance. The first miniature and complete quadrupole mass spectrometer system was introduced as the Micropole sensor. The concept featured a novel technique to assemble and operate multiple miniature quadrupoles in parallel. The short analyzer length offers a significant advantage by enabling direct mass filtering at pressures up in the 10(-2) torr range. High voltages at higher frequencies (10-20 MHz) are required for acceptable mass resolving powers. Additional trade-offs were uncovered in miniature sensors leading to designs optimized for each class of applications. Real time ray tracing of ions injected and filtered in the quadrupole field is used early in the design stage to predict the performance and reliability of the device. (C) 2001 American Society for Mass Spectrometry.
引用
收藏
页码:633 / 640
页数:8
相关论文
共 13 条
[1]  
Badman ER, 2000, J MASS SPECTROM, V35, P659, DOI 10.1002/1096-9888(200006)35:6<659::AID-JMS5>3.0.CO
[2]  
2-V
[3]  
BOUMSELLEK S, 2000, 12 ASMS SAN C FIELD, P41
[4]  
DAHL DA, 1995, EGGGS7233 ID NAT ENG
[5]  
DAWSON PH, 1976, QUADRUPOLE MASS SPEC, P125
[6]   Cold cathode field emitter array on a quadrupole mass spectrometer: Route to miniaturization [J].
Felter, TE .
JOURNAL OF VACUUM SCIENCE & TECHNOLOGY B, 1999, 17 (05) :1993-1996
[7]   High-pressure effects in miniature arrays of quadrupole analyzers for residual gas analysis from 10(-9) to 10(-2) Torr [J].
Ferran, RJ ;
Boumsellek, S .
JOURNAL OF VACUUM SCIENCE & TECHNOLOGY A, 1996, 14 (03) :1258-1265
[8]  
FERRAN RJ, 1999, Patent No. 5857890
[9]  
FERRAN RJ, 1995, Patent No. 5401962
[10]   Miniature quadrupole residual gas analyzer for process monitoring at milliTorr pressures [J].
Holkeboer, DH ;
Karandy, TL ;
Currier, FC ;
Frees, LC ;
Ellefson, RE .
JOURNAL OF VACUUM SCIENCE & TECHNOLOGY A-VACUUM SURFACES AND FILMS, 1998, 16 (03) :1157-1162