Sirolimus-eluting stents to abolish intimal hyperplasia and improve flow in porcine arteriovenous grafts - A 4-week follow-up study

被引:58
作者
Rotmans, JI
Pattynama, PMT
Verhagen, HJM
Hino, I
Velema, E
Pasterkamp, G
Stroes, ESG
机构
[1] Univ Amsterdam, Acad Med Ctr, Dept Vasc Med, NL-1105 AZ Amsterdam, Netherlands
[2] Univ Med Ctr, Expt Cardiol Lab, Utrecht, Netherlands
[3] Univ Med Ctr, Dept Vasc Surg, Utrecht, Netherlands
[4] Erasmus MC, Univ Med Ctr Rotterdam, Dept Radiol, Rotterdam, Netherlands
关键词
transplants; sirolimus; stents; hyperplasia; renal dialysis;
D O I
10.1161/01.CIR.0000159332.18585.B5
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background - The patency of arteriovenous ( AV) expanded polytetrafluoroethylene (ePTFE) hemodialysis grafts is severely compromised by intimal hyperplasia (IH) at the venous anastomosis and in the venous outflow tract. We addressed the potential of primary placement of a sirolimus-eluting stent (SES) in a validated porcine model. Methods and Results - In 25 pigs, ePTFE AV grafts were created bilaterally between the carotid artery and the jugular vein, whereupon a self-expandable nitinol stent ( 14 SESs and 11 bare-metal stents) was implanted over the venous anastomosis in 1 of the 2 grafts. After exclusion of technical failures and 1 unilateral occlusion, 16 pigs ( 9 SESs and 7 bare-metal stents) were included for further analysis. After 28 days, we measured graft flow and performed quantitative angiography. The pigs were then euthanized, and grafts with adjacent vessels were excised for histological analysis. Minimal luminal diameter was substantially larger in the SES group compared with unstented controls ( 5.9 +/- 0.2 versus 3.8 +/- 0.4 mm, respectively, P = 0.01), which was accompanied by more prominent graft flow ( SES, 1360 +/- 89 mL/min versus unstented, 861 +/- 83 mL/min, P = 0.05). IH at the venous anastomosis was 77% less in the SES group compared with unstented controls (0.44 +/- 0.05 versus 1.92 +/- 0.5 mm(2), respectively, P = 0.01), whereas IH increased markedly when bare-metal stents were used (5.7 +/- 1.4 mm(2), P = 0.05). Conclusions - SESs in the venous outflow of AV grafts significantly reduce IH and increase vessel diameter and graft flow compared with unstented grafts. These findings suggest that SESs have the potential to improve primary patency of AV grafts in hemodialysis patients.
引用
收藏
页码:1537 / 1542
页数:6
相关论文
共 29 条
[1]   Increased intimal hyperplasia in experimental vein graft stenting compared to arterial stenting: comparisons in a new rabbit model of stent injury [J].
Alp, NJ ;
West, NEJ ;
Arnold, N ;
Gunn, J ;
Banning, AP ;
Channon, KM .
CARDIOVASCULAR RESEARCH, 2002, 56 (01) :164-172
[2]   Graft flow as a predictor of thrombosis in hemodialysis grafts [J].
Bosman, PJ ;
Boereboom, FTJ ;
Eikelboom, BC ;
Koomans, HA ;
Blankestijn, PJ .
KIDNEY INTERNATIONAL, 1998, 54 (05) :1726-1730
[3]   Atherosclerotic cardiovascular disease risks in chronic hemodialysis patients [J].
Cheung, AK ;
Sarnak, MJ ;
Yan, GF ;
Dwyer, JT ;
Heyka, RJ ;
Rocco, MV ;
Teehan, BP ;
Levey, AS .
KIDNEY INTERNATIONAL, 2000, 58 (01) :353-362
[4]  
Ellis SG, 1999, CIRCULATION, V100, P799
[5]   Pathology of acute and chronic coronary stenting in humans [J].
Farb, A ;
Sangiorgi, G ;
Carter, AJ ;
Walley, VM ;
Edwards, WD ;
Schwartz, RS ;
Virmani, R .
CIRCULATION, 1999, 99 (01) :44-52
[6]   A DILEMMA FOR THE 1990S - CHOOSING APPROPRIATE EXPERIMENTAL ANIMAL-MODEL FOR THE PREVENTION OF RESTENOSIS [J].
FERRELL, M ;
FUSTER, V ;
GOLD, HK ;
CHESEBRO, JH .
CIRCULATION, 1992, 85 (04) :1630-1631
[7]   DIALYSIS ACCESS GRAFTS - ANATOMIC LOCATION OF VENOUS STENOSIS AND RESULTS ANGIOPLASTY [J].
KANTERMAN, RY ;
VESELY, TM ;
PILGRAM, TK ;
GUY, BW ;
WINDUS, DW ;
PICUS, D .
RADIOLOGY, 1995, 195 (01) :135-139
[8]   Does renal failure cause an atherosclerotic milieu in patients with end-stage renal disease? [J].
Kennedy, R ;
Case, C ;
Fathi, R ;
Johnson, D ;
Isbel, N ;
Marwick, TH .
AMERICAN JOURNAL OF MEDICINE, 2001, 110 (03) :198-204
[9]   Global burden of diabetes, 1995-2025 - Prevalence, numerical estimates, and projections [J].
King, H ;
Aubert, RE ;
Herman, WH .
DIABETES CARE, 1998, 21 (09) :1414-1431
[10]   Twenty-eight-day efficacy and phamacokinetics of the sirolimus-eluting stent [J].
Klugherz, BD ;
Llanos, G ;
Lieuallen, W ;
Kopia, GA ;
Papandreou, G ;
Narayan, P ;
Sasseen, B ;
Adelman, SJ ;
Falotico, R ;
Wilensky, RL .
CORONARY ARTERY DISEASE, 2002, 13 (03) :183-188