Spatial resolution and energy filtering of backscattered electron images in scanning electron microscopy

被引:11
作者
Merli, PG
Migliori, A
Morandi, V
Rosa, R
机构
[1] CNR, Ist LAMEL, Area Ric Bologna, I-40129 Bologna, Italy
[2] Univ Bologna, Dipartimento Sci Stat, I-40126 Bologna, Italy
关键词
scanning electron microscopy; backscattered electrons imaging; compositional contrast;
D O I
10.1016/S0304-3991(00)00132-7
中图分类号
TH742 [显微镜];
学科分类号
摘要
Numerical simulations of energy filtering effects on backscattered electron images of semiconductor multilayers are reported. The theoretical investigation has been performed for a wide range of energies, 1-40 keV, and for beam incidence angles between 90 degrees (normal incidence) and 20 degrees. Quite a general purpose of this research concerns the investigation of the optimum energy conditions and of their implications. It will be shown that the optimum energy defines an operating context suitable to ensure a compositional contrast enhancement; i.e. a minimum threshold current and a maximum resolution, without energy filtering, independent of the beam incidence angle. This optimum energy, depending on the specimen and its details, is, however, of the order of a few keV or less for specimen details having a size of the order of few nm, When the performance of the electron gun does not allow to work at low energy it is necessary to operate at an energy higher than the optimum one, the energy filtering can produce positive effects. Yet in those circumstances there is an optimum energy loss window suitable to minimise the threshold current. It spreads from 10-30%, depending on the primary energy and size of the compositional detail, for normal incidence, to a few per cent for high incidence angles and high energy. The simulation results for these last conditions are in agreement with the well-known experimental results obtained with the low-loss methods. (C) 2001 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:139 / 150
页数:12
相关论文
共 26 条
[1]   APPLICATION OF MONTE-CARLO TECHNIQUE TO THE ELECTRON-PROBE MICROANALYSIS OF TERNARY SI-B-O FILMS ON SILICON [J].
ARMIGLIATO, A ;
DESALVO, A ;
RINALDI, R ;
ROSA, R .
JOURNAL OF PHYSICS D-APPLIED PHYSICS, 1979, 12 (08) :1299-1308
[2]  
Bethe H, 1930, ANN PHYS-BERLIN, V5, P325
[3]  
BROERS AN, 1974, IITRI SEM 1, P9
[4]   THEORETICAL SIMULATION OF BACKSCATTERED ELECTRON IMAGES OF SI SIXGE1-X STRUCTURES WITH A SCANNING ELECTRON-MICROSCOPE [J].
DERICCARDIS, AC ;
MERLI, PG ;
NACUCCHI, M ;
TAPFER, L .
MIKROCHIMICA ACTA, 1994, 114 :261-266
[5]   MONTE-CARLO SIMULATION OF ELASTIC AND INELASTIC-SCATTERING OF ELECTRONS IN THIN-FILMS .2. CORE ELECTRON LOSSES [J].
DESALVO, A ;
ROSA, R .
JOURNAL OF PHYSICS D-APPLIED PHYSICS, 1987, 20 (06) :790-795
[6]  
DUNCUMB P, 1968, NBS SPECIAL PUB, V289, P133
[7]  
FRANCHI S, 1990, ELECTRON MICROS, P380
[8]  
Jiang CZ, 2000, J MICROSC-OXFORD, V198, P17, DOI 10.1046/j.1365-2818.2000.00686.x
[9]   AN EMPIRICAL STOPPING POWER RELATIONSHIP FOR LOW-ENERGY ELECTRONS [J].
JOY, DC ;
LUO, S .
SCANNING, 1989, 11 (04) :176-180
[10]   BACKSCATTERED ELECTRON CONTRAST ON CROSS-SECTIONS OF INTERFACES AND MULTILAYERS IN THE SCANNING ELECTRON-MICROSCOPE [J].
KONKOL, A ;
BOOKER, GR ;
WILSHAW, PR .
ULTRAMICROSCOPY, 1995, 58 (3-4) :233-237