Clinical evaluation of the use of the M2A patency capsule system before a capsule endoscopy procedure, in patients with known or suspected intestinal stenosis

被引:106
作者
Delvaux, M [1 ]
Ben Soussan, E
Laurent, V
Lerebours, E
Gay, G
机构
[1] CHU Nancy, Dept Internal Med & Digest Pathol, Hop Brabois Adultes, F-54511 Vandoeuvre Les Nancy, France
[2] CHU Charles Nicolle, Dept Gastroenterol, Rouen, France
[3] CHU Nancy, Hop Brabois Adultes, Dept Internal Med & Digest Pathol, F-54511 Vandoeuvre Les Nancy, France
关键词
D O I
10.1055/s-2005-870241
中图分类号
R57 [消化系及腹部疾病];
学科分类号
摘要
Background and Study Aim: The main complication of video capsule endoscopy (VCE) is the blocking of the capsule by a gastrointestinal stenosis. The "patency capsule" is a self-dissolving capsule that is the same size as the video capsule. It contains a radiofrequency identification (RFID) tag that allows it to be detected by a scanning device placed on the abdominal wall. When its passage is blocked by a stenosis, the patency capsule dissolves in 40-80 hours after ingestion. The aim of this study was to evaluate the usefulness of this system in patients with suspected intestinal stenosis but also requiring VCE. Patients and Methods: 22 patients (16 men; 46 18 years; 15 with diagnosed or suspected Crohn's disease, two with suspected intestinal tumor, and two on nonsteroidal anti-inflammatory drugs) were first investigated with an abdominal CT plus enteroclysis (entero-CT, n=15) or a small-bowel follow-through (SBFT, n=7). The patency capsule was then administered at around 09.00 to fasting patients. The presence of the patency capsule in the digestive tract was assessed 30 hours later using the hand-held scanner. When the patency capsule was detected, a plain abdominal film was obtained, and this was repeated every 24 hours until capsule expulsion. Results: At 30 hours after ingestion, the patency capsule was detected in 17 patients (72.3%): in the area of the small intestine (n=6) and in the colon (n=11). In all the patients in whom the capsule was blocked in the small intestine, the stenosis had been suspected on CT or SBFT. In three patients, the delay in progression of the patency capsule led to cancellation of the VCE procedure. In three patients, the patency capsule induced a symptomatic intestinal occlusion, which resolved spontaneously in one and required emergency surgery in two. Although these two patients would have been obliged to undergo operation for their primary disease in any case, the procedure had to be brought forward. In one patient, the capsule passed through the stenosis without symptoms. All patients in whom the progression of the patency capsule was delayed had an established or suspected diagnosis of Crohn ' s disease. Conclusions: The current technical development of the patency capsule limits its use in clinical practice, as it did not detect ste noses undiagnosed by CT or SBFT. The start of dissolution at 40 hours after ingestion is too slow to prevent episodes of intestinal occlusion. Patients with Crohn ' s disease are most likely to be at risk of blockage of progression of the capsule and should benefit from a CT investigation before VCE. However, a careful interview eliciting the patient ' s medical history and symptoms remains the most useful indicator with regard to suspicion of an intestinal stenosis.
引用
收藏
页码:801 / 807
页数:7
相关论文
共 18 条
  • [1] Barkin Jamie S, 2004, Gastrointest Endosc Clin N Am, V14, P61, DOI 10.1016/j.giec.2003.10.016
  • [2] Wireless Capsule Endoscopy requiring surgical intervention: The world's experience
    Barkin, JS
    Friedman, S
    [J]. AMERICAN JOURNAL OF GASTROENTEROLOGY, 2002, 97 (09) : S298 - S298
  • [3] The utility of computed tomography in acute small bowel obstruction
    Burkill, GJC
    Bell, JRG
    Healy, JC
    [J]. CLINICAL RADIOLOGY, 2001, 56 (05) : 350 - 359
  • [4] Costamagna G, 2004, GASTROINTEST ENDOSC, V59, pAB145
  • [5] Wireless capsule video endoscopy is a superior diagnostic tool in comparison to barium follow-through and computerized tomography in patients with suspected Crohn's disease
    Eliakim, R
    Fischer, D
    Suissa, A
    Yassin, K
    Katz, D
    Guttman, N
    Migdal, M
    [J]. EUROPEAN JOURNAL OF GASTROENTEROLOGY & HEPATOLOGY, 2003, 15 (04) : 363 - 367
  • [6] The first prospective controlled trial comparing wireless capsule endoscopy with push enteroscopy in chronic gastrointestinal bleeding
    Ell, C
    Remke, S
    May, A
    Helou, L
    Henrich, R
    Mayer, G
    [J]. ENDOSCOPY, 2002, 34 (09) : 685 - 689
  • [7] Temporary intestinal occlusion induced by a "patency capsule" in a patient with Crohn's disease
    Gay, G
    Delvaux, M
    Laurent, V
    Reibel, N
    Regent, D
    Grosdidier, G
    Roche, JF
    [J]. ENDOSCOPY, 2005, 37 (02) : 174 - 177
  • [8] The role of video capsule endoscopy in the diagnosis of digestive diseases: a review of current possibilities
    Gay, G
    Delvaux, M
    Rey, JF
    [J]. ENDOSCOPY, 2004, 36 (10) : 913 - 920
  • [9] Herrerias JM, 2003, ENDOSCOPY, V35, P564
  • [10] Wireless capsule endoscopy
    Iddan, G
    Meron, G
    Glukhovsky, A
    Swain, P
    [J]. NATURE, 2000, 405 (6785) : 417 - 417