Fast, low noise preamplifiers for pulsed opto-thermal radiometry

被引:1
作者
Berg, EP [1 ]
O'Driscoll, D [1 ]
Imhof, RE [1 ]
机构
[1] S Bank Univ, Sch EEIE, London SE1 0AA, England
来源
PHOTOACOUSTIC AND PHOTOTHERMAL PHENOMENA: TENTH INTERNATIONAL CONFERENCE | 1999年 / 463卷
关键词
D O I
10.1063/1.58189
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The generally poor performance of commercial preamplifiers for Mercury Cadmium Telluride detectors in high speed transient capture applications have led us to produce our own designs. We summarise methods for predicting performance, illustrated with a design for a photoconductive detector of 4x4 mm sensitive area, giving a noise-equivalent temperature sensitivity of 1.5 mK at 1.2 MHz bandwidth, with high immunity from synchronous radio frequency interference.
引用
收藏
页码:141 / 143
页数:3
相关论文
共 2 条
[1]  
*AN DEV, 1992, AD797 DATA SHEET
[2]  
Dereniak E., 1996, INFRARED DETECTORS S