Picogram measurement of volatile n-alkanes (n-hexane through n-dodecane) in blood using solid-phase microextraction to assess nonoccupational petroleum-based fuel exposure

被引:32
作者
Chambers, David M. [1 ]
Blount, Benjamin C. [1 ]
McElprang, David O. [1 ]
Waterhouse, Michael G. [1 ]
Morrow, John C. [1 ]
机构
[1] Ctr Dis Control & Prevent, Div Sci Lab, Natl Ctr Environm Hlth, Atlanta, GA 30341 USA
关键词
D O I
10.1021/ac800065d
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
We describe here a new method for the analysis of alkanes (n-hexane, n-heptane, n-octane, n-nonane, n-decane, n-undecane, and n-dodecane) in blood using headspace solid-phase microextraction gas chromatography/mass spectrometry. Ibis method is used to measure picogram per milliliter levels of n-alkanes in blood that may result from nonoccupational exposure to alkanes and other volatile nonpolar compounds from common sources such as petroleum-based fuel. This alkane signature is useful in distinguishing typical fuel biomarkers (e.g., benzene and toluene) from other confounding exposure sources such as cigarette smoke. Development of this method required special attention to sample handling as alkanes are not highly soluble in aqueous matrixes and exist as ubiquitous compounds found in many laboratory materials and the environment. In particular, significant n-hexane contamination (similar to 0.4 ng/mL) occurred from collecting blood samples in vacutainers. This residue was removed by boiling the vacutainer stoppers in methanol followed by vacuum baking. For all the alkanes, the calculated accuracy demonstrated for the water-based standards ranged from 3.3% to 17% as deduced from the difference of the lowest and middle standards from the curve fit. Quality control data among runs over a 10 month period were found to vary from 14% to -29%, with a few exceptions. The resulting quantification limits for n-hexane through n-decane ranged from 0.069 to 0.132 ng/mL. In the analysis of 1200 blood samples from people with no known occupational exposure, median blood levels for all n-alkanes were below these quantification limits. n-Hexane levels above the method detection limit were, however, found in 1.3% of the samples.
引用
收藏
页码:4666 / 4674
页数:9
相关论文
共 30 条
[1]  
[Anonymous], 1987, Quality Assurance of Chemical Measurements
[2]   Using the blood concentration of 2,5-dimethylfuran as a marker for smoking [J].
Ashley, DL ;
Bonin, MA ;
Hamar, B ;
McGeehin, M .
INTERNATIONAL ARCHIVES OF OCCUPATIONAL AND ENVIRONMENTAL HEALTH, 1996, 68 (03) :183-187
[3]   Quantification of 31 volatile organic compounds in whole blood using solid-phase microextraction and gas chromatography-mass spectrometry [J].
Blount, BC ;
Kobelski, RJ ;
McElprang, DO ;
Ashley, DL ;
Morrow, JC ;
Chambers, DM ;
Cardinali, FL .
JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2006, 832 (02) :292-301
[4]   UBIQUITOUS POLLUTION BY NORMAL-HEXANE AND REFERENCE BIOLOGICAL LEVELS IN THE GENERAL-POPULATION [J].
BRUGNONE, F ;
MARANELLI, G ;
ROMEO, L ;
GIULIARI, C ;
GOBBI, M ;
MALESANI, F ;
BASSI, G ;
ALEXOPOULOS, C .
INTERNATIONAL ARCHIVES OF OCCUPATIONAL AND ENVIRONMENTAL HEALTH, 1991, 63 (03) :157-160
[5]   Potential health effects of gasoline and its constituents: A review of current literature (1990-1997) on toxicological data [J].
Caprino, L ;
Togna, GI .
ENVIRONMENTAL HEALTH PERSPECTIVES, 1998, 106 (03) :115-125
[6]   PRODUCTION OF BLANK WATER FOR THE ANALYSIS OF VOLATILE ORGANIC-COMPOUNDS IN HUMAN BLOOD AT THE LOW PARTS-PER-TRILLION LEVEL [J].
CARDINALI, FL ;
MCCRAW, JM ;
ASHLEY, DL ;
BONIN, MA .
JOURNAL OF CHROMATOGRAPHIC SCIENCE, 1994, 32 (01) :41-45
[7]   TREATMENT OF VACUTAINERS FOR USE IN THE ANALYSIS OF VOLATILE ORGANIC-COMPOUNDS IN HUMAN BLOOD AT THE LOW PARTS-PER-TRILLION LEVEL [J].
CARDINALI, FL ;
MCCRAW, JM ;
ASHLEY, DL ;
BONIN, M ;
WOOTEN, J .
JOURNAL OF CHROMATOGRAPHIC SCIENCE, 1995, 33 (10) :557-560
[8]   An improved approach for accurate quantitation of benzene, toluene, ethylbenzene, xylene, and styrene in blood [J].
Chambers, David M. ;
McElprang, David O. ;
Waterhouse, Michael G. ;
Blount, Benjamin C. .
ANALYTICAL CHEMISTRY, 2006, 78 (15) :5375-5383
[9]   Identification and elimination of polysiloxane curing agent interference encountered in the quantification of low-picogram per milliliter methyl tert-butyl ether in blood by solid-phase microextraction headspace analysis [J].
Chambers, DM ;
McElprang, DO ;
Mauldin, JP ;
Hughes, TM ;
Blount, BC .
ANALYTICAL CHEMISTRY, 2005, 77 (09) :2912-2919
[10]   Determination of benzene and associated volatile compounds in mainstream cigarette smoke [J].
Darrall, KG ;
Figgins, JA ;
Brown, RD ;
Phillips, GF .
ANALYST, 1998, 123 (05) :1095-1101