Determination of 25-hydroxyvitamin D in human plasma using high-performance liquid chromatography-tandem mass spectrometry
被引:108
作者:
Tsugawa, N
[1
]
Suhara, Y
[1
]
Kamao, M
[1
]
Okano, T
[1
]
机构:
[1] Kobe Pharmaceut Univ, Dept Hyg Sci, Higashinada Ku, Kobe, Hyogo 6588558, Japan
关键词:
D O I:
10.1021/ac048249c
中图分类号:
O65 [分析化学];
学科分类号:
070302 ;
081704 ;
摘要:
We report here the development of a precise and sensitive method to determine 25-hydroxyvitamin D (25-OH-D-2/-D-3) in human plasma using high-perfomance liquid chromatography-tandem mass-mass spectrometry with atomospheric pressure chemical ionization (LC-APCI-MS/ MS). The method involves the use of deuterated 25-OH-D-3 as an internal standard compound for 25-OH-D-2/-D-3, which was synthesized in our laboratory, and the selection of a precursor and product ion with a MS/MS multiple reaction monitoring method. The average intraassay and interassay variation values (relative standard deviation) were 5.7 and 2.5%, respectively, for 25-OH-D-3 and 4.5 and 5.1%, respectively, for 25-OH-D-2. The average spiked recoveries from authentic compounds added to normal human plasma samples for 25-OH-D-3 and 25-OH-D-2 were 103.8 and 98.8%, respectively. Mean plasma concentrations of 25-OH-D-3 and 25-OH-D-2 in healthy subjects were 20.5 and 0.4 ng/mL, respectively. We conclude that this novel LC-APCI-MS/MS method would be useful for the evaluation of the vitamin D status in postmenopausal women and elderly subjects and provide useful information in the diagnosis of vitamin D insufficiency/deficiency, as well as for the treatment and prevention of osteoporosis with vitamin D.
引用
收藏
页码:3001 / 3007
页数:7