Aortic neck attachment failure and the AneuRx graft: Incidence, treatment options, and early results

被引:14
作者
Azizzadeh, A [1 ]
Sanchez, LA [1 ]
Rubin, BG [1 ]
Parodi, JC [1 ]
Godshall, CJ [1 ]
Geraghty, PJ [1 ]
Choi, ET [1 ]
Flye, MW [1 ]
Curci, JA [1 ]
Sicard, GA [1 ]
机构
[1] Washington Univ, Sch Med, Dept Surg, Vasc Surg Sect, St Louis, MO 63110 USA
关键词
D O I
10.1007/s10016-005-4619-4
中图分类号
R61 [外科手术学];
学科分类号
摘要
Some investigators have reported that proximal attachment failure is a long-term complication of endovascular abdominal aortic aneurysm repair (EVAR) with the AneuRx (Medtronic, Santa Rosa, CA) device. We evaluated the need for an intervention in patients with suboptimal proximal fixation as well as the feasibility and early success of a variety of treatment strategies. From October 1999 to October 2003, we performed 365 EVARs using the AneuRx graft. At a mean follow-up of 23.7 +/- 14.8 months, 20 patients (5.5%) with suboptimal outcomes (14 with a type I endoleak, one with a type III endoleak, and 5 with an inadequate seal zone < 1 cm) were considered for treatment. Characteristics of each patient''s aortic neck anatomy that could be associated with proximal attachment failure were evaluated. Eighteen patients (90%) underwent successful treatment (9 AneuRx cuffs, 6 Talent cuffs, 5 aortic stents, one redo endograft, and two surgical conversions) without major perioperative complications, one patient had a persistent type I endoleak despite endovascular treatment, and one patient refused treatment, ultimately leading to aneurysm rupture. There have been no further endoleaks or graft migrations noted since the secondary intervention at a mean follow-up of 13.9 +/- 11.8 months. In our experience, proximal attachment failure associated with the AneuRx graft is relatively uncommon and usually associated with unfavorable neck anatomy. Despite this, most cases are treatable by endovascular means. Long-term follow-up is needed to assess the ultimate frequency of these combined device reconstructions.
引用
收藏
页码:516 / 521
页数:6
相关论文
共 8 条
[1]   Device migration after endoluminal abdominal aortic aneurysm repair: Analysis of 113 cases with a minimum follow-up period of 2 years [J].
Cao, P ;
Verzini, F ;
Zannetti, S ;
De Rango, P ;
Parlani, G ;
Lupattelli, L ;
Maselli, A .
JOURNAL OF VASCULAR SURGERY, 2002, 35 (02) :229-235
[2]   Identifying and grading factors that modify the outcome of endovascular aortic aneurysm repair [J].
Chaikof, EL ;
Fillinger, MF ;
Matsumura, JS ;
Rutherford, RB ;
White, GH ;
Blankensteijn, JD ;
Bernhard, VM ;
Harris, PL ;
Kent, KC ;
May, J ;
Veith, FJ ;
Zarins, CK .
JOURNAL OF VASCULAR SURGERY, 2002, 35 (05) :1061-1066
[3]   Endograft migration one to four years after endovascular abdominal aortic aneurysm repair with the AneuRx device: A cautionary note [J].
Conners, MS ;
Sternbergh, WC ;
Carter, G ;
Tonnessen, BH ;
Yoselevitz, M ;
Money, SR .
JOURNAL OF VASCULAR SURGERY, 2002, 36 (03) :476-482
[4]  
*MEDTR VASC, 2004, CLIN UPD ANEURX STEN, V2
[5]   Nature, frequency, and predictors of secondary procedures after endovascular repair of abdominal aortic aneurysm [J].
Sampram, ESK ;
Karafa, MT ;
Mascha, EJ ;
Clair, DG ;
Greenberg, RK ;
Lyden, SP ;
O'Hara, PJ ;
Sarac, TP ;
Srivastava, SD ;
Butler, B ;
Ouriel, K .
JOURNAL OF VASCULAR SURGERY, 2003, 37 (05) :930-936
[6]   Aortic neck angulation predicts adverse outcome with endovascular abdominal aortic aneurysm repair [J].
Sternbergh, WC ;
Carter, G ;
York, JW ;
Yoselevitz, M ;
Money, SR .
JOURNAL OF VASCULAR SURGERY, 2002, 35 (03) :482-486
[7]   Stent graft migration after endovascular aneurysm repair: Importance of proximal fixation [J].
Zarins, CK ;
Bloch, DA ;
Crabtree, T ;
Matsumoto, AH ;
White, RA ;
Fogarty, TJ .
JOURNAL OF VASCULAR SURGERY, 2003, 38 (06) :1264-1272
[8]   The US AneuRx Clinical Trial: 6-year clinical update 2002 [J].
Zarins, CK .
JOURNAL OF VASCULAR SURGERY, 2003, 37 (04) :904-908