Benefit of higher closed-loop bandwidths in ocular adaptive optics

被引:67
作者
Diaz-Santana, L
Torti, C
Munro, I
Gasson, P
Dainty, C
机构
[1] City Univ London, Dept Optometry & Visual Sci, Appl Vis Res Ctr, London EC1V 0HB, England
[2] Univ London Imperial Coll Sci Technol & Med, Blackett Lab, Photon Grp, London SW7 2BW, England
[3] Natl Univ Ireland, Dept Expt Phys, Galway, Ireland
来源
OPTICS EXPRESS | 2003年 / 11卷 / 20期
关键词
D O I
10.1364/OE.11.002597
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
We present an ocular adaptive optics system with a wavefront sampling rate of 240 Hz and maximum recorded closed-loop bandwidth close to 25 Hz, but with typical performances around 10 Hz. The measured bandwidth depended on the specific system configuration and the particular subject tested. An analysis of the system performance as a function of achieved bandwidth showed consistently higher Strehl ratios for higher closed-loop bandwidths. This may be attributed to a combination of limitations on the available technology and the dynamics of ocular aberrations. We observed dynamic behaviour with a maximum frequency content around 30 Hz. (C) 2003 Optical Society of America.
引用
收藏
页码:2597 / 2605
页数:9
相关论文
共 14 条
[1]  
Diaz Santana Haro L, 1999, Opt Lett, V24, P61
[2]   Effects of retinal scattering in the ocular double-pass process [J].
Diaz-Santana, L ;
Dainty, JC .
JOURNAL OF THE OPTICAL SOCIETY OF AMERICA A-OPTICS IMAGE SCIENCE AND VISION, 2001, 18 (07) :1437-1444
[3]   Use of a microelectromechanical mirror for adaptive optics in the human eye [J].
Doble, N ;
Yoon, G ;
Chen, L ;
Bierden, P ;
Singer, B ;
Olivier, S ;
Williams, DR .
OPTICS LETTERS, 2002, 27 (17) :1537-1539
[4]   Closed-loop adaptive optics in the human eye [J].
Fernández, EJ ;
Iglesias, I ;
Artal, P .
OPTICS LETTERS, 2001, 26 (10) :746-748
[5]  
GLANC M, 2002, THESIS U PAIS SUD
[6]   Improvement in retinal image quality with dynamic correction of the eye's aberrations [J].
Hofer, H ;
Chen, L ;
Yoon, GY ;
Singer, B ;
Yamauchi, Y ;
Williams, DR .
OPTICS EXPRESS, 2001, 8 (11) :631-643
[7]   Dynamics of the eye's wave aberration [J].
Hofer, H ;
Artal, P ;
Singer, B ;
Aragón, JL ;
Williams, DR .
JOURNAL OF THE OPTICAL SOCIETY OF AMERICA A-OPTICS IMAGE SCIENCE AND VISION, 2001, 18 (03) :497-506
[8]   Measurement of fine dynamic changes of the corneal topography by use of interferometry [J].
Kasprzak, HT ;
Jaronski, JT .
INTERFEROMETRY XI: APPLICATIONS, 2002, 4778 :169-176
[9]  
Larichev AV, 2001, QUANTUM ELECTRON+, V31, P1108, DOI 10.1070/QE2001v031n12ABEH002105
[10]   Supernormal vision and high-resolution retinal imaging through adaptive optics [J].
Liang, JZ ;
Williams, DR ;
Miller, DT .
JOURNAL OF THE OPTICAL SOCIETY OF AMERICA A-OPTICS IMAGE SCIENCE AND VISION, 1997, 14 (11) :2884-2892