We report measurements with a micromachined mechanical biosensor that inherently suppresses background effects by producing a differential signal with respect to a reference. The sensor comprises two adjacent cantilevers with interdigitated fingers between them that allow interferometric detection of the differential, i.e., relative bending. We demonstrate that differential detection can efficiently suppress unspecific chemical effects that result in cantilever bending. We show that the differential deflection noise is up to an order of magnitude lower than the absolute deflection noise in the low-frequency range of 0.0003-1 Hz, where many types of biologically relevant reactions occur. We also demonstrate the sensor's applicability to biological receptor-ligand systems by reporting experimental results on direct differential detection of biotin-streptavidin binding. (C) 2003 American Institute of Physics.
引用
收藏
页码:1659 / 1661
页数:3
相关论文
共 11 条
[1]   Label-free protein assay based on a nanomechanical cantilever array [J].
Arntz, Y ;
Seelig, JD ;
Lang, HP ;
Zhang, J ;
Hunziker, P ;
Ramseyer, JP ;
Meyer, E ;
Hegner, M ;
Gerber, C .
NANOTECHNOLOGY, 2003, 14 (01) :86-90
[2]   A sensitive method to measure changes in the surface stress of solids [J].
Butt, HJ .
JOURNAL OF COLLOID AND INTERFACE SCIENCE, 1996, 180 (01) :251-260
[3]   Translating biomolecular recognition into nanomechanics [J].
Fritz, J ;
Baller, MK ;
Lang, HP ;
Rothuizen, H ;
Vettiger, P ;
Meyer, E ;
Güntherodt, HJ ;
Gerber, C ;
Gimzewski, JK .
SCIENCE, 2000, 288 (5464) :316-318
[4]   ION-SENSING DEVICES WITH SILICON-NITRIDE AND BOROSILICATE GLASS INSULATORS [J].
HARAME, DL ;
BOUSSE, LJ ;
SHOTT, JD ;
MEINDL, JD .
IEEE TRANSACTIONS ON ELECTRON DEVICES, 1987, 34 (08) :1700-1707
[5]   Detection of pH variation using modified microcantilever sensors [J].
Ji, HF ;
Hansen, KM ;
Hu, Z ;
Thundat, T .
SENSORS AND ACTUATORS B-CHEMICAL, 2001, 72 (03) :233-238
[6]   Optimization and performance of high-resolution micro-optomechanical thermal sensors [J].
Lai, J ;
Perazzo, T ;
Shi, Z ;
Majumdar, A .
SENSORS AND ACTUATORS A-PHYSICAL, 1997, 58 (02) :113-119
[7]   Sub-10 cm3 interferometric accelerometer with nano-g resolution [J].
Loh, NC ;
Schmidt, MA ;
Manalis, SR .
JOURNAL OF MICROELECTROMECHANICAL SYSTEMS, 2002, 11 (03) :182-187
[8]   Interdigital cantilevers for atomic force microscopy [J].
Manalis, SR ;
Minne, SC ;
Atalar, A ;
Quate, CF .
APPLIED PHYSICS LETTERS, 1996, 69 (25) :3944-3946
[9]   Fabrication and characterization of a micromechanical sensor for differential detection of nanoscale motions [J].
Savran, CA ;
Sparks, AW ;
Sihler, J ;
Li, J ;
Wu, WC ;
Berlin, DE ;
Burg, TP ;
Fritz, J ;
Schmidt, MA ;
Manalis, SR .
JOURNAL OF MICROELECTROMECHANICAL SYSTEMS, 2002, 11 (06) :703-708
[10]   Bioassay of prostate-specific antigen (PSA) using microcantilevers [J].
Wu, GH ;
Datar, RH ;
Hansen, KM ;
Thundat, T ;
Cote, RJ ;
Majumdar, A .
NATURE BIOTECHNOLOGY, 2001, 19 (09) :856-860