Multicentre evaluation of KONE Optima analysis system

被引:2
作者
Ferrero, C
Modenese, A
Bonini, PA
Ceriotti, F
Carobene, A
Franzin, M
Guder, WG
Hofmann, W
Puukka, M
Laitinen, H
Makarovsky, V
Gaidukova, N
机构
[1] San Raffaele Sci Inst, Lab Anal, I-20132 Milan, Italy
[2] Stadt Krankenhaus Munchen Bogenhausen, Inst Klin Chem, Munchen, Germany
[3] Oulu Univ Hosp Lab, Oulu, Finland
[4] Cent Clin Hosp Russian Federat, Moscow, Russia
[5] KONE Instruments Corp, Espoo, Finland
关键词
chemistry analyzer; ion selective electrodes; laboratory management; precision;
D O I
10.1515/CCLM.1998.080
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Optima from KONE Instruments Corporation is a new selective laboratory analyzer for turbidimetric, colorimetric and ion selective electrode measurements. Overall analytical performance of Optima and reagents provided by KONE was evaluated according to ECCLS guidelines, in a multicentre study involving four different laboratories, including substrates (cholesterol, high-density lipoprotein-cholesterol, creatinine), specific proteins (transferrin, IgG), enzyme activities (gamma-glutamyltransferase, alanine aminotransferase) and electrolytes (sodium, potassium, chloride). The results obtained attest good precision of the system for all the analytes tested: the range of within-run CV is 0.5 %-4.3 %, and range of between-day CV % is 0.8 %-7.9 % (median of four laboratories). Except for total cholesterol (5 % overestimation compared to the reference method) accuracy of measurement is adequate. Creatinine and uric acid assays were subjects of interference (defined as deviation > 10 % from target value) by bilirubin and haemoglobin respectively.
引用
收藏
页码:475 / 484
页数:10
相关论文
共 9 条
[1]   APPLICATION OF STATISTICAL PROCEDURES IN ANALYTICAL INSTRUMENT TESTING [J].
BABLOK, W ;
PASSING, H .
JOURNAL OF AUTOMATIC CHEMISTRY, 1985, 7 (02) :74-79
[2]  
BONINI P, 1992, EUR J CLIN CHEM CLIN, V30, P881
[3]   REVISED SCHEME FOR EVALUATION OF AUTOMATIC INSTRUMENTS FOR USE IN CLINICAL CHEMISTRY [J].
BROUGHTON, PM ;
GOWENLOCK, AH ;
MCCORMACK, JJ ;
NEILL, DW .
ANNALS OF CLINICAL BIOCHEMISTRY, 1974, 11 (NOV) :207-218
[4]  
Duncan I.W., 1982, PROCEDURE PROPOSED C
[5]  
*ECCLS, 1986, 2 ECCLS
[6]  
FRASER CG, 1992, EUR J CLIN CHEM CLIN, V30, P311
[7]  
GLICK MR, 1986, CLIN CHEM, V32, P470
[8]  
*LAB STAND PAN NAT, 1988, CLIN CHEM, V34, P193
[9]  
SHAW LM, 1983, J CLIN CHEM CLIN BIO, V21, P633