Continuous Glucose Monitoring-A Study of the Enlite Sensor During Hypo- and Hyperbaric Conditions

被引:17
作者
Adolfsson, Peter [1 ]
Ornhagen, Hans [2 ]
Eriksson, Bengt M. [3 ]
Cooper, Ken [4 ]
Jendle, Johan [5 ,6 ]
机构
[1] Univ Gothenburg, Goteborg Pediat Growth Res Ctr, Dept Pediat, Inst Clin Sci,Sahlgrenska Acad, Gothenburg, Sweden
[2] Swedish Sports Diving Federat, Farsta, Sweden
[3] Karolinska Univ Hosp, Dept Anesthesiol, Solna, Sweden
[4] Medtron Diabet Sensor R&D, Northridge, CA USA
[5] Karlstad Hosp, Endocrine & Diabet Ctr, Karlstad, Sweden
[6] Orebro Univ Hosp, Fac Med & Hlth, Orebro, Sweden
关键词
NOCTURNAL HYPOGLYCEMIA; PERFORMANCE; ACCURACY; EXERCISE;
D O I
10.1089/dia.2011.0284
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background: The performance and accuracy of the Enlite (TM) (Medtronic, Inc., Northridge, CA) sensor may be affected by microbubble formation at the electrode surface during hypo- and hyperbaric conditions. The effects of acute pressure changes and of prewetting of sensors were investigated. Materials and Methods: On Day 1, 24 sensors were inserted on the right side of the abdomen and back in one healthy individual; 12 were prewetted with saline solution, and 12 were inserted dry. On Day 2, this procedure was repeated on the left side. All sensors were attached to an iPro continuous glucose monitoring (CGM) recorder. Hypobaric and hyperbaric tests were conducted in a pressure chamber, with each test lasting 105 min. Plasma glucose values were obtained at 5-min intervals with a HemoCue (R) (Angelholm, Sweden) model 201 glucose analyzer for comparison with sensor glucose values. Results: Ninety percent of the CGM systems operated during the tests. The mean absolute relative difference was lower during hyperbaric than hypobaric conditions (6.7% vs. 14.9%, P<0.001). Sensor sensitivity was slightly decreased (P<0.05) during hypobaric but not during hyperbaric conditions. Clarke Error Grid Analysis showed that 100% of the values were found in the A+B region. No differences were found between prewetted and dry sensors. Conclusions: The Enlite sensor performed adequately during acute pressure changes and was more accurate during hyperbaric than hypobaric conditions. Prewetting the sensors did not improve accuracy. Further studies on type 1 diabetes subjects are needed under various pressure conditions.
引用
收藏
页码:527 / 532
页数:6
相关论文
共 19 条
[1]  
Adolfsson Peter, 2008, J Diabetes Sci Technol, V2, P778
[2]   Accuracy and Reliability of Continuous Glucose Monitoring in Individuals with Type 1 Diabetes During Recreational Diving [J].
Adolfsson, Peter ;
Ornhagen, Hans ;
Jendle, Johan .
DIABETES TECHNOLOGY & THERAPEUTICS, 2009, 11 (08) :493-497
[3]  
Amtzen A, 2008, NORWEGIAN DIVING TRE
[4]   Effect of Continuous Glucose Monitoring on Hypoglycemia in Type 1 Diabetes [J].
Battelino, Tadej ;
Phillip, Moshe ;
Bratina, Natasa ;
Nimri, Revital ;
Oskarsson, Per ;
Bolinder, Jan .
DIABETES CARE, 2011, 34 (04) :795-800
[6]   The Effect of Continuous Glucose Monitoring in Well-Controlled Type 1 Diabetes [J].
Beck, Roy W. .
DIABETES CARE, 2009, 32 (08) :1378-1383
[7]   Safety of recreational scuba diving in type 1 diabetic patients: The Deep Monitoring programme [J].
Bonomo, M. ;
Cairoli, R. ;
Verde, G. ;
Morelli, L. ;
Moreo, A. ;
Delle Grottaglie, M. ;
Brambilla, M. C. ;
Meneghini, E. ;
Aghemo, P. ;
Corigliano, G. ;
Marroni, A. .
DIABETES & METABOLISM, 2009, 35 (02) :101-107
[8]  
Davey Raymond J, 2010, J Diabetes Sci Technol, V4, P1457
[9]  
Fink Kenneth S, 2002, Diabetes Technol Ther, V4, P627, DOI 10.1089/152091502320798259
[10]   Influence of simulated altitude ore the performance of five blood glucose meters [J].
Gautier, JF ;
Bigard, AX ;
Douce, P ;
Duvallet, A ;
Cathelineau, G .
DIABETES CARE, 1996, 19 (12) :1430-1433