Human precision of operating a digitizing board: Implications for electrocardiogram measurements

被引:19
作者
Malik, M [1 ]
Bradford, A [1 ]
机构
[1] St George Hosp, Sch Med, Dept Cardiol Sci, London SW17 0RE, England
来源
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY | 1998年 / 21卷 / 08期
关键词
digitizing board; measurement precision; reproducibility; electrocardiogram; QT dispersion;
D O I
10.1111/j.1540-8159.1998.tb00255.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
To investigate the precision achieved by human measurement on a digitizing board, 100 healthy volunteers (46 women, mean age 36 +/- 12 years) were asked to measure 15 times an artificial pattern composed of 15 points. A high precision digitizing board (programmed to the technical accuracy of +/- 50 mu m) was used, and mean and maximum errors in measuring the same distance repeatedly and relocalizing the same point repeatedly were obtained for each volunteer. A median mean and maxim um error of 0.2 mm and 1.0 mm were found for repeated distance measurement. When simulating QT dispersion measurement (measuring the same distance 12 times), median value of 20 ms was obtained for ECGs of 25 mm/s paper speed. The study concludes that human precision of operating a digitizing board is rather poor. A recommendation is given to use either a computer screen for manual measurement of ECGs or 50 provide an operator of the digitizing board with an immediate feedback of the precision and measurement stability achieved so that erroneous measurement can be actively rejected.
引用
收藏
页码:1656 / 1662
页数:7
相关论文
共 8 条
[1]  
Bhullar HK, 1992, COMPUT CARDIOL, P533
[2]  
DAY CP, 1990, BRIT HEART J, V63, P342
[3]  
HNATKOVA K, 1996, EUR JPCE, V6, P113
[4]   SHORT-TERM AND LONG-TERM REPRODUCIBILITY OF QT, QTC, AND QT DISPERSION MEASUREMENT IN HEALTHY-SUBJECTS [J].
KAUTZNER, J ;
YI, G ;
CAMM, AJ ;
MALIK, M .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1994, 17 (05) :928-937
[5]  
Kautzner J, 1996, ANN NONINVAS ELECTRO, V1, P363
[6]   VARIABILITY OF QT DISPERSION MEASUREMENTS IN THE SURFACE ELECTROCARDIOGRAM IN PATIENTS WITH ACUTE MYOCARDIAL-INFARCTION AND IN NORMAL SUBJECTS [J].
VANDELOO, A ;
ARENDTS, W ;
HOHNLOSER, SH .
AMERICAN JOURNAL OF CARDIOLOGY, 1994, 74 (11) :1113-1118
[7]   ELECTROCARDIOGRAPHIC INDEXES OF DISPERSION OF VENTRICULAR REPOLARIZATION - AN ISOLATED HEART VALIDATION-STUDY [J].
ZABEL, M ;
PORTNOY, S ;
FRANZ, MR .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1995, 25 (03) :746-752
[8]  
ZABEL M, IN PRESS CIRCULATION