An improved outdoor calibration procedure for broadband ultraviolet radiometers

被引:18
作者
Cancilllo, ML [1 ]
Serrano, A
Antón, M
García, JA
Vilaplana, JM
de la Morena, B
机构
[1] Univ Extremadura, Dept Fis, E-06071 Badajoz, Spain
[2] ESAt, INTA, Dept Observac Tierra Teledetecc & Atomsfera, Huelva, Spain
关键词
D O I
10.1562/2005-01-12-RA-412R.1
中图分类号
Q5 [生物化学]; Q7 [分子生物学];
学科分类号
071010 ; 081704 ;
摘要
This article aims at improving the broadband ultraviolet radiometer's calibration methodology. For this goal, three broadband radiometers are calibrated against a spectrophotometer of reference. Three different one-step calibration models are tested: ratio, first order and second order. The latter is proposed in order to adequately reproduce the high dependence on the solar zenith angle shown by the other two models and, therefore, to improve the calibration performance at high solar elevations. The proposed new second-order model requires no additional information and, thus, keeps the operational character of the one-step methodology. The models are compared in terms of their root mean square error and the most qualified is subsequently validated by comparing its predictions with the spectrophotometer measurements within an independent validation data subset. Results show that the best calibration is achieved by the second-order model, with a mean bias error and mean absolute bias error lower than 2.2 and 6.7%, respectively.
引用
收藏
页码:860 / 865
页数:6
相关论文
共 30 条
[1]  
[Anonymous], CIE J
[2]   Assessment of four methods to estimate surface UV radiation using satellite data, by comparison with ground measurements from four stations in Europe [J].
Arola, A ;
Kalliskota, S ;
den Outer, PN ;
Edvardsen, K ;
Hansen, G ;
Koskela, T ;
Martin, TJ ;
Matthijsen, J ;
Meerkoetter, R ;
Peeters, P ;
Seckmeyer, G ;
Simon, PC ;
Slaper, H ;
Taalas, P ;
Verdebout, J .
JOURNAL OF GEOPHYSICAL RESEARCH-ATMOSPHERES, 2002, 107 (D16) :11-1
[3]  
BAIS A, 1999, 1051 WMO GAW LAPCOST
[4]  
Bodhaine BA, 1998, J ATMOS OCEAN TECH, V15, P916, DOI 10.1175/1520-0426(1998)015<0916:CBUIOA>2.0.CO
[5]  
2
[6]   UV-B ROBERTSON-BERGER METER CHARACTERIZATION AND FIELD CALIBRATION [J].
GRAINGER, RG ;
BASHER, RE ;
MCKENZIE, RL .
APPLIED OPTICS, 1993, 32 (03) :343-349
[7]   Solar UV measurements with Robertson-Berger type instruments: influence of the detector's internal humidity status [J].
Huber, M ;
Blumthaler, M ;
Schreder, J .
AGRICULTURAL AND FOREST METEOROLOGY, 2003, 120 (1-4) :39-43
[8]   Effect of ambient temperature on Robertson-Berger-type erythernal dosimeters [J].
Huber, M ;
Blumthaler, M ;
Schreder, J ;
Bais, A ;
Topaloglou, C .
APPLIED OPTICS, 2002, 41 (21) :4273-4277
[9]   Comparison of Brewer spectrophotometer ultraviolet data from similar latitudes in the northern and southern hemisphere [J].
Kimlin, MG ;
Sabburg, JM ;
Parisi, A ;
Meltzer, RS .
JOURNAL OF ATMOSPHERIC AND SOLAR-TERRESTRIAL PHYSICS, 2003, 65 (16-18) :1401-1410
[10]  
KYLLING A, 2001, LIBRADTRAN SOFTWARE