Sparse random ultrasound phased array for focal surgery

被引:113
作者
Goss, SA
Frizzell, LA
Kouzmanoff, JT
Barich, JM
Yang, JM
机构
[1] LABTHERM TECHNOL INC,CHAMPAIGN,IL 61820
[2] UNIV ILLINOIS,DEPT ELECT & COMP ENGN,URBANA,IL 61801
基金
美国国家卫生研究院;
关键词
D O I
10.1109/58.542054
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
Ultrasound phased arrays offer several advantages over single focused transducer technology, enabling electronically programmable synthesis of focal size and shape, as well as position, While phased arrays have been employed for medical diagnostic and therapeutic (hyperthermia) applications, there remain fundamental Problems associated with their use for surgery. These problems stem largely from the Small size of each array element dictated by the wavelength employed at surgical application frequencies (2-4 MHz), the array aperture size required for the desired focal characteristics, and the number of array elements and electronic drive channels required to provide RF energy to the entire array, The present work involves the theoretical and experimental examination of novel ultrasound phased arrays consisting of array elements larger than one wavelength, minimizing the humber of elements in an aperture through a combination of geometric focusing, directive beams, and sparse random placement of array elements, for tissue ablation applications. A hexagonally packed array consisting of 108 I-mm-diameter circular elements mounted on a spherical shell was modeled theoretically and a prototype array was constructed to examine the feasibility of sparse random array configurations for focal surgery, A randomly selected subset of elements of the prototype test array (64 of 108 available channels) was driven at 2.1 MHz with a 64-channel digitally controlled RF drive system. The performance of the prototype array was evaluated by comparing field data obtained from theoretical modeling to that obtained experimentally via hydrophone scanning, The results of that comparison, along with total acoustic power measurements, suggest that the use of sparse random phased arrays for focal surgery is feasible, and that the nature of array packing is an important determinant to observed performance.
引用
收藏
页码:1111 / 1121
页数:11
相关论文
共 33 条
[1]  
CHAPELON JY, P 1993 IEEE ULTR S, P1211
[2]   EXPERIMENTAL EVALUATION OF A PROTOTYPE CYLINDRICAL SECTION ULTRASOUND HYPERTHERMIA PHASED-ARRAY APPLICATOR [J].
EBBINI, ES ;
CAIN, CA .
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 1991, 38 (05) :510-520
[3]   ABILITY OF HIGH-INTENSITY ULTRASOUND TO ABLATE HUMAN ATHEROSCLEROTIC PLAQUES AND MINIMIZE DEBRIS SIZE [J].
ERNST, A ;
SCHENK, EA ;
GRACEWSKI, SM ;
WOODLOCK, TJ ;
MURANT, FG ;
ALLIGER, H ;
MELTZER, RS .
AMERICAN JOURNAL OF CARDIOLOGY, 1991, 68 (02) :242-246
[4]  
FAN X, 1994, MED PHYS, V22, P297
[5]  
FOSTER RS, 1993, EUR UROL, V23, P29
[6]  
Frizzell L. A., 1987, Proceedings of the Ninth Annual Conference of the IEEE Engineering in Medicine and Biology Society (Cat. No.87CH2513-0), P1642
[7]   THRESHOLD DOSAGES FOR DAMAGE TO MAMMALIAN LIVER BY HIGH-INTENSITY FOCUSED ULTRASOUND [J].
FRIZZELL, LA .
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 1988, 35 (05) :578-581
[8]   THRESHOLDS FOR FOCAL ULTRASONIC LESIONS IN RABBIT KIDNEY, LIVER AND TESTICLE [J].
FRIZZELL, LA ;
LINKE, CA ;
CARSTENSEN, EL ;
FRIDD, CW .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1977, 24 (04) :393-396
[9]  
FRIZZELL LA, 1985 IEEE ULTR S P, P930
[10]  
FRY F J, 1978, Ultrasound in Medicine and Biology, V4, P337, DOI 10.1016/0301-5629(78)90022-4