Accuracy of cortical and trabecular bone measurements with peripheral quantitative computed tomography (pQCT)

被引:119
作者
Augat, P
Gordon, CL
Lang, TF
Iida, H
Genant, HK
机构
[1] Univ Calif San Francisco, Dept Radiol, Osteoporosis & Arthrit Res Grp, San Francisco, CA 94143 USA
[2] Univ Calif San Francisco, Dept Orthoped Surg, Hand & Microvasc Res Lab, San Francisco, CA 94143 USA
关键词
D O I
10.1088/0031-9155/43/10/015
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
In order to assess the accuracy of peripheral QCT (Stratec XCT 960) we analysed scans of the European Forearm Phantom and another phantom consisting of K2HPO4 encased in aluminium tubes to simulate cortical walls. Additionally 14 cadaveric forearm specimen scans were compared to CT scans acquired on a GE9800Q. The accuracy for density assessment of the European Forearm Phantom was better than 3%. A small increase in density was observed with increasing thickness of the aluminium wall (10% for each mm). Density measurements within the wall were confounded by limited spatial resolution. For a thickness of less than 4 mm, the density within the wall was underestimated by up to 40%. The measurement of mineral content was not influenced by this effect and showed an accuracy error of less than 6%. The agreement of density measurements on the different CT systems was very strong (R-2 > 0.96; RMSE<6.2%). Our findings suggest that the Stratec pQCT scanner very accurately measures volumetric trabecular and total bone mineral densities at the distal radius while the assessment of cortical density is associated with considerable inaccuracies due to limited spatial resolution.
引用
收藏
页码:2873 / 2883
页数:11
相关论文
共 30 条
[1]   Assessment of bone mineral at appendicular sites in females with fractures of the proximal femur [J].
Augat, P ;
Fan, B ;
Lane, NE ;
Lang, TF ;
LeHir, P ;
Lu, Y ;
Uffmann, M ;
Genant, HK .
BONE, 1998, 22 (04) :395-402
[2]  
AUGAT P, 1997, OSTEOPOROSIS INT, V7, P269
[3]   FOREARM BMD AS MEASURED BY PERIPHERAL QUANTITATIVE COMPUTED-TOMOGRAPHY (PQCT) IN A GERMAN REFERENCE POPULATION [J].
BUTZ, S ;
WUSTER, C ;
SCHEIDTNAVE, C ;
GOTZ, M ;
ZIEGLER, R .
OSTEOPOROSIS INTERNATIONAL, 1994, 4 (04) :179-184
[4]   FIXED-RATIO BETWEEN RADIAL CORTICAL VOLUME AND DENSITY MEASURED BY PERIPHERAL QUANTITATIVE COMPUTED-TOMOGRAPHY (PQCT) REGARDLESS OF AGE AND SEX [J].
FUJII, Y ;
MIYAUCHI, A ;
TAKAGI, Y ;
GOTO, B ;
FUJITA, T .
CALCIFIED TISSUE INTERNATIONAL, 1995, 56 (06) :586-588
[5]   Effect of aging on trabecular and compact bone components of proximal and ultradistal radius [J].
Gatti, D ;
Rossini, M ;
Zamberlan, N ;
Braga, V ;
Fracassi, E ;
Adami, S .
OSTEOPOROSIS INTERNATIONAL, 1996, 6 (05) :355-360
[6]   IMPACT OF MARROW FAT ON ACCURACY OF QUANTITATIVE CT [J].
GLUER, CC ;
GENANT, HK .
JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1989, 13 (06) :1023-1035
[7]   In vivo assessment of trabecular bone structure at the distal radius from high-resolution computed tomography images [J].
Gordon, CL ;
Webber, CE ;
Adachi, JD ;
Christoforou, N .
PHYSICS IN MEDICINE AND BIOLOGY, 1996, 41 (03) :495-508
[8]   Comparisons of noninvasive bone mineral measurements in assessing age-related loss, fracture discrimination, and diagnostic classification [J].
Grampp, S ;
Genant, HK ;
Mathur, A ;
Lang, P ;
Jergas, M ;
Takada, M ;
Gluer, CC ;
Lu, Y ;
Chavez, M .
JOURNAL OF BONE AND MINERAL RESEARCH, 1997, 12 (05) :697-711
[9]  
GRAMPP S, 1995, J BONE MINER RES, V10, P1566
[10]   CORRECTION OF SCATTER IN COMPUTED-TOMOGRAPHY IMAGES OF BONE [J].
HANGARTNER, TN .
MEDICAL PHYSICS, 1987, 14 (03) :335-340