Detection of consciousness by electroencephalogram and auditory evoked potentials

被引:40
作者
Schneider, G
Hollweck, R
Ningler, M
Stockmanns, G
Kochs, EF
机构
[1] Tech Univ Munich, Klinikum Isar, Dept Anesthesiol, D-81675 Munich, Germany
[2] Univ Duisburg Gesamthsch, Inst Informat Technol, Duisburg, Germany
关键词
BISPECTRAL INDEX; MONITORING DEPTH; AWARENESS; ANESTHESIA; PROPOFOL; RESPONSES; ENTROPY;
D O I
10.1097/00000542-200511000-00006
中图分类号
R614 [麻醉学];
学科分类号
100217 [麻醉学];
摘要
Background: A set of electroencephalographic and auditory evoked potential (AEP) parameters should be identified that allows lows separation of consciousness from unconsciousness (reflected by responsiveness/unresponsiveness to command). Methods: Forty unpremedicated patients received anesthesia with remifentanil and either sevoflurane or propofol. With remifentanil infusion (0.2 mu g(.)kg(-1.)min(-1)), patients were asked every 30 s to squeeze the investigator's hand. Sevoflurane or propofol was given until loss of consciousness. After intubation, propofol or sevoflurane was stopped until patients followed the command (return of consciousness). Thereafter, propofol or sevoflurane was started again (loss of consciousness), and surgery was performed. Return of consciousness was observed after surgery. The electroencephalogram and AEP from immediately before and after the transitions were selected. Logistic regression was calculated to identify models for the separation between consciousness and unconsciousness. For the top 10 models, 1,000-fold cross-validation was performed. Backward variable selection was applied to identify a minimal model. Prediction probability was calculated. The digitized electroencephalogram was replayed, and the Bispectral Index was measured and accordingly analyzed. Results: The best full model (prediction probability 0.89) contained 15 AEP and 4 electroencephalographic parameters. The best minimal model (prediction probability 0.87) contained 2 AEP and 2 electroencephalographic parameters (median frequency of the amplitude spectrum from 8 - 30 Hz and approximate entropy). The prediction probability of the Bispectral Index was 0.737. Conclusions: A combination of electroencephalographic and AEP parameters can he used to differentiate between consciousness and unconsciousness even in a very challenging data set. The minimal model contains a combination of AEP and electroencephalographic parameters and has a higher prediction probability than Bispectral Index for the separation between consciousness and unconsciousness.
引用
收藏
页码:934 / 943
页数:10
相关论文
共 25 条
[1]
A SIMPLE STUDY OF AWARENESS AND DREAMING DURING ANAESTHESIA [J].
BRICE, DD ;
HETHERINGTON, RR ;
UTTING, JE .
BRITISH JOURNAL OF ANAESTHESIA, 1970, 42 (06) :535-+
[2]
Shannon entropy applied to the measurement of the electroencephalographic effects of desflurane [J].
Bruhn, J ;
Lehmann, LE ;
Röpcke, H ;
Bouillon, TW ;
Hoeft, A .
ANESTHESIOLOGY, 2001, 95 (01) :30-35
[3]
Approximate entropy as an electroencephalographic measure of anesthetic drug effect during desflurane anesthesia [J].
Bruhn, J ;
Röpcke, H ;
Hoeft, A .
ANESTHESIOLOGY, 2000, 92 (03) :715-726
[4]
A tutorial on Support Vector Machines for pattern recognition [J].
Burges, CJC .
DATA MINING AND KNOWLEDGE DISCOVERY, 1998, 2 (02) :121-167
[5]
Awareness and the EEG power spectrum: analysis of frequencies [J].
Dressler, O ;
Schneider, G ;
Stockmanns, G ;
Kochs, EF .
BRITISH JOURNAL OF ANAESTHESIA, 2004, 93 (06) :806-809
[6]
Monitoring depth of anesthesia - With emphasis on the application of the bispectral index and the middle latency auditory evoked response to the prevention of recall [J].
Drummond, JC .
ANESTHESIOLOGY, 2000, 93 (03) :876-882
[7]
Comparison of bispectral EEG analysis and auditory evoked potentials for monitoring depth of anaesthesia during propofol anaesthesia [J].
Gajraj, RJ ;
Doi, M ;
Mantzaridis, H ;
Kenny, GNC .
BRITISH JOURNAL OF ANAESTHESIA, 1999, 82 (05) :672-678
[8]
HEARING AND MEMORY IN ANESTHETIZED PATIENTS [J].
JONES, JG ;
KONIECZKO, K .
BRITISH MEDICAL JOURNAL, 1986, 292 (6531) :1291-1293
[9]
MONITORING EVOKED-POTENTIALS DURING SURGERY TO ASSESS THE LEVEL OF ANESTHESIA [J].
JORDAN, C ;
WELLER, C ;
THORNTON, C ;
NEWTON, DEF .
JOURNAL OF MEDICAL ENGINEERING & TECHNOLOGY, 1995, 19 (2-3) :77-79
[10]
Monitors of depth of anestbesia, quo vadis? [J].
Kalkman, CJ ;
Drummond, JC .
ANESTHESIOLOGY, 2002, 96 (04) :784-787