Overloading of the second-dimension column in comprehensive two-dimensional gas chromatography

被引:36
作者
Harynuk, J
Górecki, T
de Zeeuw, J
机构
[1] Univ Waterloo, Dept Chem, Waterloo, ON N2L 3G1, Canada
[2] Varian Inc, NL-4330 EA Middelburg, Netherlands
关键词
comprehensive two-dimensional gas chromatography; GC x GC; overloading; column diameter;
D O I
10.1016/j.chroma.2004.11.049
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Comprehensive two-dimensional gas chromatography (GC x GC) is based on a coupling of two GC columns of different characteristics by means of a device that allows portions of the effluent from the primary column to be injected onto the second dimension column for an additional separation. The time available for the separation in the second-dimension column is very short. Thus, this separation should be very efficient. The vast majority of GC x GC practitioners use very narrow bore columns for the second dimension. While this approach is justified in principle, if peaks in the second dimension overload this column, its peak capacity is severely reduced. A series of second-dimension columns of varying internal diameters, but similar phase ratios, were used to study these effects. The results indicate that 250 mu m columns often provide comparable second dimension peak widths to 100 mu m columns, while at the same time being less prone to overloading, indicating that they may often be a better choice than smaller diameter columns in the second dimension of GC x GC systems. (c) 2004 Elsevier B.V. All rights reserved.
引用
收藏
页码:21 / 27
页数:7
相关论文
共 7 条
[1]   Fast comprehensive two-dimensional gas chromatography (GC x GC) using 50-μm ID second-dimension columns [J].
Adahchour, M ;
Tasöz, A ;
Beens, J ;
Vreuls, RJJ ;
Batenburg, AM ;
Brinkman, UAT .
JOURNAL OF SEPARATION SCIENCE, 2003, 26 (9-10) :753-760
[2]   Comprehensive two-dimensional gas chromatography:: a powerful and versatile analytical tool [J].
Dallüge, J ;
Beens, J ;
Brinkman, UAT .
JOURNAL OF CHROMATOGRAPHY A, 2003, 1000 (1-2) :69-108
[3]   The evolution of comprehensive two-dimensional gas chromatography (GC x GC) [J].
Górecki, T ;
Harynuk, J ;
Panic, O .
JOURNAL OF SEPARATION SCIENCE, 2004, 27 (5-6) :359-379
[4]   New liquid nitrogen cryogenic modulator for comprehensive two-dimensional gas chromatography [J].
Harynuk, J ;
Górecki, T .
JOURNAL OF CHROMATOGRAPHY A, 2003, 1019 (1-2) :53-63
[5]   Effect of sampling rate on resolution in comprehensive two-dimensional liquid chromatography [J].
Murphy, RE ;
Schure, MR ;
Foley, JP .
ANALYTICAL CHEMISTRY, 1998, 70 (08) :1585-1594
[6]  
Ong R, 2001, J SEP SCI, V24, P367, DOI 10.1002/1615-9314(20010501)24:5<367::AID-JSSC367>3.0.CO
[7]  
2-U