Numerical analysis and experimental estimation of a low-leakage injection technique for capillary electrophoresis

被引:53
作者
Fu, LM
Lin, CH [1 ]
机构
[1] Natl Sun Yat Sen Univ, Dept Mech & Electromech Engn, Kaohsiung 804, Taiwan
[2] Natl Pingtung Univ Sci & Technol, Grad Inst Mat Engn, Pingtung 912, Taiwan
关键词
D O I
10.1021/ac034667q
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
This paper presents an experimental and numerical investigation into the use of low-leakage injection techniques to deliver sample plugs within electrophoresis microchips. The study addresses the principal material transport mechanisms such as electrokinetic migration, fluid flow, and diffusion and gives detail analyses to the double-L injection technique, which employs electrokinetic manipulations to avoid sample leakage within the microchip. Electrical potential contour, velocity vector, and streamline distribution in the micro CE chip are successfully developed. Experimental and numerical testing results show the double-L injection technique is capable of reducing sample leakage within cross-form microfluidic chips. The current study confirms the double-L injection technique has an exciting potential for use in high-quality, high-throughput chemical analysis applications and in many other applications throughout the field of micro total analysis systems.
引用
收藏
页码:5790 / 5796
页数:7
相关论文
共 45 条
[1]  
Alarie JP, 2001, ELECTROPHORESIS, V22, P312, DOI 10.1002/1522-2683(200101)22:2<312::AID-ELPS312>3.0.CO
[2]  
2-3
[3]  
Alarie JP, 2000, ELECTROPHORESIS, V21, P100, DOI 10.1002/(SICI)1522-2683(20000101)21:1<100::AID-ELPS100>3.3.CO
[4]  
2-G
[5]   Micro total analysis systems. 2. Analytical standard operations and applications [J].
Auroux, PA ;
Iossifidis, D ;
Reyes, DR ;
Manz, A .
ANALYTICAL CHEMISTRY, 2002, 74 (12) :2637-2652
[6]   A chip-based electrophoresis system with electrochemical detection and hydrodynamic injection [J].
Backofen, U ;
Matysik, FM ;
Lunte, CE .
ANALYTICAL CHEMISTRY, 2002, 74 (16) :4054-4059
[7]   Finite element simulation of an electroosmotic-driven flow division at a T-junction of microscale dimensions [J].
Bianchi, F ;
Ferrigno, A ;
Girault, HH .
ANALYTICAL CHEMISTRY, 2000, 72 (09) :1987-1993
[8]   Microchip injection and separation anomalies due to pressure effects [J].
Crabtree, HJ ;
Cheong, ECS ;
Tilroe, DA ;
Backhouse, CJ .
ANALYTICAL CHEMISTRY, 2001, 73 (17) :4079-4086
[9]   Electroosmotically induced hydraulic pumping on microchips: Differential ion transport [J].
Culbertson, CT ;
Ramsey, RS ;
Ramsey, JM .
ANALYTICAL CHEMISTRY, 2000, 72 (10) :2285-2291
[10]   Analytical solution of combined eloectoossmotic/pressure driven flows in two-dimensional straight channels: Finite debye layer effects [J].
Dutta, P ;
Beskok, A .
ANALYTICAL CHEMISTRY, 2001, 73 (09) :1979-1986