Portable XRF analysis of occupational air filter samples from different workplaces using different sampler: final results, summary and conclusion

被引:12
作者
Harper, Martin
Pacolay, Bruce
Hintz, Patrick
Bartly, David L.
Slaven, James E.
Andrew, Michael E.
机构
[1] NIOSH, Exposure Assessment Branch, Hlth Effects Lab Div, Morgantown, WV 26505 USA
[2] NIOSH, Spokane Res Lab, Cincinnati, OH 45226 USA
[3] NIOSH, Cincinnati, OH 45226 USA
[4] NIOSH, Biostat & Epidemiol Branch, Hlth Effects Lab Div, Morgantown, WV 26505 USA
来源
JOURNAL OF ENVIRONMENTAL MONITORING | 2007年 / 9卷 / 11期
关键词
D O I
10.1039/b710591f
中图分类号
O65 [分析化学];
学科分类号
070302 [分析化学]; 081704 [应用化学];
摘要
This paper concludes a five-year program on research into the use of a portable X-ray fluorescence (XRF) analyzer for analyzing lead in air sampling filters from different industrial environments, including mining, manufacturing and recycling. The results from four of these environments have already been reported. The results from two additional metal processes are presented here. At both of these sites, lead was a minor component of the total airborne metals and interferences from other elements were minimal. Nevertheless, only results from the three sites where lead was the most abundant metal were used in the overall calculation of method accuracy. The XRF analyzer was used to interrogate the filters, which were then subjected to acid digestion and analysis by inductively-coupled plasma optical-emission spectroscopy (ICP-OES). The filter samples were collected using different filter-holders or "samplers'' where the size (diameter), depth and homogeneity of aerosol deposit varied from sampler to sampler. The aerosol collection effeciencies of the samplers were expected to differ, especially for larger particles. The distribution of particles once having entered the sampler was also expected to differ between samplers. Samplers were paired to allow the between-sampler variability to be addressed, and, in some cases, internal sampler wall deposits were evaluated and compared to the filter catch. It was found, rather surprisingly, that analysis of the filter deposits (by ICP- OES) of all the samplers gave equivalent results. It was also found that deposits on some of the sampler walls, which in some protocols are considered part of the sample, could be significant in comparison to the filter deposit. If it is concluded that wall-deposits should be analyzed, then XRF analysis of the filter can only give a minimum estimate of the concentration. Techniques for the statistical analysis of field data were also developed as part of this program and have been reported elsewhere. The results, based on data from the three workplaces where lead was the major element present in the samples, are summarized here. A limit of detection and a limit of quantitation are provided. Analysis of some samples using a second analyzer with a different X-ray source technology indicated reasonable agreement for some metals (but this was not evaluated for lead). Provided it is only necessary to analyze the filters, most personal samplers will provide acceptable results when used with portable XRF analysis for lead around applicable limit values.
引用
收藏
页码:1263 / 1270
页数:8
相关论文
共 16 条
[1]
BARTLEY DL, 2007, IN PRESS J OCCUP ENV
[2]
Demange Martine, 2002, Appl Occup Environ Hyg, V17, P200, DOI 10.1080/104732202753438289
[3]
Evaluation of two portable lead-monitoring methods at mining sites [J].
Drake, PL ;
Lawryk, NJ ;
Ashley, K ;
Sussell, AL ;
Hazelwood, KJ ;
Song, RG .
JOURNAL OF HAZARDOUS MATERIALS, 2003, 102 (01) :29-38
[4]
ELLER P, 1994, PUBLICATION NATL I O
[5]
ELLER P, 1998, PUBLICATION NATL I O
[6]
ELSKAMP CJ, 2004, OSHA MANUAL ANAL MET
[7]
A comparison of X-ray fluorescence and wet chemical analysis for lead on air filters from different personal samplers used in a secondary lead smelter/solder manufacturer [J].
Harper, M ;
Pacolay, B .
JOURNAL OF ENVIRONMENTAL MONITORING, 2006, 8 (01) :140-146
[8]
A comparison of portable XRF and ICP-OES analysis for lead on air filter samples from a lead ore concentrator mill and a lead-acid battery recycler [J].
Harper, M ;
Pacolay, B ;
Hintz, P ;
Andrew, ME .
JOURNAL OF ENVIRONMENTAL MONITORING, 2006, 8 (03) :384-392
[9]
A comparison of X-ray fluorescence and wet chemical analysis for lead on air filters from different personal samplers used in a bronze foundry [J].
Harper, M ;
Pacolay, B ;
Andrew, ME .
JOURNAL OF ENVIRONMENTAL MONITORING, 2005, 7 (06) :592-597
[10]
A comparison of X-ray fluorescence and wet chemical analysis of air filter samples from a scrap lead smelting operation [J].
Harper, M ;
Hallmark, TS ;
Andrew, ME ;
Bird, AJ .