Multi-step dip-spin coating manufacturing system for silicone cardiovascular membrane fabrication with prescribed compliance

被引:14
作者
Cortez, Miguel A.
Quintana, Rolando
Wicker, Ryan B. [1 ]
机构
[1] Univ Texas San Antonio, El Paso, TX 79968 USA
[2] Univ Texas San Antonio, San Antonio, TX 78285 USA
关键词
cardiovascular system modeling; compliant vessel fabrication; dip-spin coating; rotational curing; silicone membranes;
D O I
10.1007/s00170-006-0649-5
中图分类号
TP [自动化技术、计算机技术];
学科分类号
0812 ;
摘要
A multi-step dip-spin coating procedure by which synthetic vessels of virtually the entire cardiovascular system can be constructed from medical imaging data was developed to consistently and accurately fabricate silicone membranes with prescribed compliance for in vitro hemodynamic experimentation and medical device development. Experimental results showed that the three factors tested (silicone mixture viscosity, mold diameter, and number of dips in solution) each had a statistically significant effect on the resulting wall thickness of the fabricated silicone membranes. A setup was designed and constructed to test the compliance of the silicone membranes, and a prediction model showed that the most important factors for compliance were the mean pressure used to pressurize the membranes, the membrane wall thickness, and the difference in membrane diameter between the two test pressures. The manufacturing process was repeatable with a process standard deviation of 0.002753 cm for an average wall thickness of 0.063576 cm (4.33%).
引用
收藏
页码:667 / 679
页数:13
相关论文
共 19 条
[1]   EFFECT OF COMPLIANCE MISMATCH ON VASCULAR GRAFT PATENCY [J].
ABBOTT, WM ;
MEGERMAN, J ;
HASSON, JE ;
LITALIEN, G ;
WARNOCK, DF .
JOURNAL OF VASCULAR SURGERY, 1987, 5 (02) :376-382
[2]  
Ahumada GG, 1987, CARDIOVASCULAR PATHO
[3]  
[Anonymous], QUALITY PLANNING ANA
[4]   Fabrication of multi-layer composite hollow fiber membranes for gas separation [J].
Chung, TS ;
Shieh, JJ ;
Lau, WWY ;
Srinivasan, MP ;
Paul, DR .
JOURNAL OF MEMBRANE SCIENCE, 1999, 152 (02) :211-225
[5]  
CORTEZ MA, 2002, FABRICATION SILICONE
[6]  
*DUP DOW EL, 2001, NEOPR LAT DIPP GOOD
[7]  
HERGER JW, 1998, CARDIOLOGY
[8]   Particle image velocimetry measurements in complex geometries [J].
Hopkins, LM ;
Kelly, JT ;
Wexler, AS ;
Prasad, AK .
EXPERIMENTS IN FLUIDS, 2000, 29 (01) :91-95
[9]   Optical filters on linear halogen-lamps prepared by dip-coating [J].
Kostlin, H ;
Frank, G ;
Hebbinghaus, G ;
Auding, H ;
Denissen, K .
JOURNAL OF NON-CRYSTALLINE SOLIDS, 1997, 218 :347-353
[10]  
Montgomery DC., 1997, Design and analysis of experiments, V4