Electric field tomography: setup for single-channel measurements

被引:3
作者
Korjenevsky, A. V. [1 ]
Tuykin, T. S. [1 ]
机构
[1] Russian Acad Sci, Inst Radioengn & Elect, Moscow 125009, Russia
关键词
electric field tomography; electrical impedance imaging; bioimpedance; contactless measurement; experimental implementation;
D O I
10.1088/0967-3334/28/7/S21
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
Results of experiments on electric field tomography, which were carried out to test the theoretical model suggested previously, are described in this paper. The theoretical foundation of the electric field tomography and substantiation of the possibility of its biomedical application are briefly presented. An experimental setup and test samples used for measurements are described. The results of experiments are presented; their comparison with the results of numerical simulation is given. Further steps towards increasing measurement accuracy and implementation of multi-channel tomographic systems are suggested.
引用
收藏
页码:S279 / S289
页数:11
相关论文
共 12 条
[1]   Preliminary static EIT images of the thorax in health and disease [J].
Cherepenin, V ;
Karpov, A ;
Korjenevsky, A ;
Kornienko, V ;
Kultiasov, Y ;
Mazaletskaya, A ;
Mazourov, D .
PHYSIOLOGICAL MEASUREMENT, 2002, 23 (01) :33-41
[2]   Three-dimensional EIT imaging of breast tissues: System design and clinical testing [J].
Cherepenin, VA ;
Karpov, AY ;
Korjenevsky, AV ;
Kornienko, VN ;
Kultiasov, YS ;
Ochapkin, MB ;
Trochanova, OV ;
Meister, JD .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 2002, 21 (06) :662-667
[3]  
GABRIEL C, 1996, ALOETR19960037 KINGS
[4]   Magnetic induction tomography [J].
Griffiths, H .
MEASUREMENT SCIENCE AND TECHNOLOGY, 2001, 12 (08) :1126-1131
[5]   Magnetic induction tomography: experimental realization [J].
Korjenevsky, A ;
Cherepenin, V ;
Sapetsky, S .
PHYSIOLOGICAL MEASUREMENT, 2000, 21 (01) :89-94
[6]   Maxwell-Wagner relaxation in electrical imaging [J].
Korjenevsky, AV .
PHYSIOLOGICAL MEASUREMENT, 2005, 26 (02) :S101-S110
[7]   Electric field tomography for contactless imaging of resistivity in biomedical applications [J].
Korjenevsky, AV .
PHYSIOLOGICAL MEASUREMENT, 2004, 25 (01) :391-401
[8]   Recent developments and industrial research applications of capacitance tomography [J].
Reinecke, N ;
Mewes, D .
MEASUREMENT SCIENCE AND TECHNOLOGY, 1996, 7 (03) :233-246
[9]   A multifrequency magnetic induction tomography system using planar gradiometers: data collection and calibration [J].
Rosell-Ferrer, J. ;
Merwa, R. ;
Brunner, P. ;
Scharfetter, H. .
PHYSIOLOGICAL MEASUREMENT, 2006, 27 (05) :S271-S280
[10]   ELECTRICAL CAPACITANCE TOMOGRAPHY FOR FLOW IMAGING - SYSTEM MODEL FOR DEVELOPMENT OF IMAGE-RECONSTRUCTION ALGORITHMS AND DESIGN OF PRIMARY SENSORS [J].
XIE, CG ;
HUANG, SM ;
HOYLE, BS ;
THORN, R ;
LENN, C ;
SNOWDEN, D ;
BECK, MS .
IEE PROCEEDINGS-G CIRCUITS DEVICES AND SYSTEMS, 1992, 139 (01) :89-98