Characterization of diffuse reflectance fiber probe sampling on moving solids using a Fourier transform near-infrared spectrometer

被引:23
作者
Berntsson, O [1 ]
Danielsson, LG
Folestad, S
机构
[1] Royal Inst Technol, Div Analyt Chem, SE-10044 Stockholm, Sweden
[2] AstraZeneca R&D Molndal, Pharmaceut & Analyt R&D, SE-43183 Molndal, Sweden
关键词
FT-NIR spectrometer; interferograms; fiber-optic probes;
D O I
10.1016/S0003-2670(00)01313-1
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A characterization of the effect of sample movement on spectral response using an FT-NIR spectrometer is presented. The data were evaluated by means of both multivariate data analysis and visual inspection of interferograms and spectra. A fiber-optic probe was used to collect spectra from model samples such as paper, plastic and fine and coarse powders. During controlled movement at velocities between 0.046 and 0.73 m s(-1), it was found that sample movement causes features not normally present in the interferograms. However, it is shown that after transformation to frequency domain single-beam spectra these artifacts are present well outside the near infrared spectral range (10,000-4000 cm(-1)). The implication of these results is that FT-NIR is a suitable technique for process analysis of samples moving with low or moderate velocities. (C) 2001 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:125 / 131
页数:7
相关论文
共 23 条
[1]   Quantitative analysis of film coating in a fluidized bed process by in line NIR spectrometry and multivariate batch calibration [J].
Andersson, M ;
Folestad, S ;
Gottfries, J ;
Johansson, MO ;
Josefson, M ;
Wahlund, KG .
ANALYTICAL CHEMISTRY, 2000, 72 (09) :2099-2108
[2]   Monitoring of a film coating process for tablets using near infrared reflectance spectrometry [J].
Andersson, M ;
Josefson, M ;
Langkilde, FW ;
Wahlund, KG .
JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 1999, 20 (1-2) :27-37
[3]  
ANDERSSON M, 2001, IN PRESS ANAL CHIM A
[4]  
[Anonymous], 1989, MULTIVARIATE CALIBRA
[5]   Effective sample size in diffuse reflectance near-IR spectrometry [J].
Berntsson, O ;
Burger, T ;
Folestad, S ;
Danielsson, LG ;
Kuhn, J ;
Fricke, J .
ANALYTICAL CHEMISTRY, 1999, 71 (03) :617-623
[6]   Estimation of effective sample size when analysing powders with diffuse reflectance near-infrared spectrometry [J].
Berntsson, O ;
Danielsson, LG ;
Folestad, S .
ANALYTICA CHIMICA ACTA, 1998, 364 (1-3) :243-251
[7]   Quantitative determination of the scattering and absorption coefficients from diffuse reflectance and transmittance measurements: Application to pharmaceutical powders [J].
Burger, T ;
Kuhn, J ;
Caps, R ;
Fricke, J .
APPLIED SPECTROSCOPY, 1997, 51 (03) :309-317
[8]  
Burger T., 1998, J NEAR INFRARED SPEC, V6, P33
[9]  
CIURCZAK EW, 1997, PHARM TECHNOL, V18
[10]   On-line monitoring of powder blending with near-infrared spectroscopy [J].
De Maesschalck, R ;
Sanchez, FC ;
Massart, DL ;
Doherty, P ;
Hailey, P .
APPLIED SPECTROSCOPY, 1998, 52 (05) :725-731