Relationship between pacemaker dependency and the effect of pacing mode on cardiovascular outcomes

被引:70
作者
Tang, ASL
Roberts, RS
Kerr, C
Gillis, AM
Green, MS
Talajic, M
Yusuf, S
Abdollah, H
Gent, M
Connolly, SJ
机构
[1] Univ Ottawa, Inst Heart, Ottawa, ON K1Y 4W7, Canada
[2] McMaster Univ, Hamilton, ON, Canada
[3] Univ British Columbia, Vancouver, BC V5Z 1M9, Canada
[4] Univ Calgary, Calgary, AB, Canada
[5] Montreal Heart Inst, Montreal, PQ H1T 1C8, Canada
[6] Queens Univ, Kingston, ON, Canada
关键词
pacemakers; trials; mortality;
D O I
10.1161/01.CIR.103.25.3081
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background-A recently completed trial, the Canadian Trial of Physiological Pacing (CTOPP), showed that physiological pacing did not significantly reduce mortality, stroke, or heart failure hospitalization, but it did show that atrial fibrillation occurred less frequently in patients with physiological pacing. Many pacemaker patients experience only transient bradyarrhythmias with an adequate unpaced heart rate (UHR) and are not pacemaker-dependent. The purpose of the present analysis was to determine if pacemaker-dependent patients have an increased benefit from physiological pacing compared with non-pacemaker-dependent patients. Methods and Results-Of 2568 patients included in the CTOPP trial, 2244 patients had a pacemaker dependency test performed at the first follow-up visit. The yearly event rate of cardiovascular death or stroke steadily increased with decreasing UHR in the ventricular pacing group, but it remained constant in the physiological pacing group. When the patients were subdivided to UHR less than or equal to 60 bpm or >60 bpm, there was an interaction between pacing mode treatment and UHR subgroup. The Kaplan-Meier plot confirmed a physiological pacing advantage only in the UHR less than or equal to 60 bpm subgroup. This differential effect was also present for the outcomes of cardiovascular death and total mortality. Conclusions-This study demonstrated that UHR at first follow-up has an important influence on how pacing mode selection affects cardiovascular death and total mortality. Pacemaker-dependent patients with low UHR will probably be paced frequently and will likely benefit from physiological pacing. In contrast, non-pacemaker-dependent patients will likely be paced infrequently and may not benefit from physiological pacing.
引用
收藏
页码:3081 / 3085
页数:5
相关论文
共 14 条
[1]   Pacing in sick sinus syndrome - Need for a prospective, randomized trial comparing atrial with dual chamber pacing [J].
Andersen, HR ;
Nielsen, JC .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1998, 21 (06) :1175-1179
[2]   PROSPECTIVE RANDOMIZED TRIAL OF ATRIAL VERSUS VENTRICULAR PACING IN SICK-SINUS SYNDROME [J].
ANDERSEN, HR ;
THUESEN, L ;
BAGGER, JP ;
VESTERLUND, T ;
THOMSEN, PEB .
LANCET, 1994, 344 (8936) :1523-1528
[3]   Long-term follow-up of patients from a randomised trial of atrial versus ventricular pacing for sick-sinus syndrome [J].
Andersen, HR ;
Nielsen, JC ;
Thomsen, PEB ;
Thuesen, L ;
Mortensen, PT ;
Vesterlund, T ;
Pedersen, AK .
LANCET, 1997, 350 (9086) :1210-1216
[4]   Dual-chamber versus ventricular pacing - Critical appraisal of current data [J].
Connolly, SJ ;
Kerr, C ;
Gent, M ;
Yasuf, S .
CIRCULATION, 1996, 94 (03) :578-583
[5]   Effects of physiologic pacing versus ventricular pacing on the risk of stroke and death due to cardiovascular causes [J].
Connolly, SJ ;
Kerr, CR ;
Gent, M ;
Roberts, RS ;
Yusuf, S ;
Gillis, AM ;
Sami, MH ;
Talajic, M ;
Tang, ASL ;
Klein, GJ ;
Lau, C ;
Newman, DM .
NEW ENGLAND JOURNAL OF MEDICINE, 2000, 342 (19) :1385-1391
[6]   IDENTIFICATION OF PACEMAKER DEPENDENT PATIENTS BY SERIAL DECREMENTAL RATE INHIBITION [J].
CRICK, JCP ;
ROKAS, S ;
SOWTON, E .
EUROPEAN HEART JOURNAL, 1985, 6 (10) :891-896
[7]  
Gillis AM, 1998, CAN J CARDIOL, V14, P1115
[8]  
JANOSIK DL, 1995, CLIN CARDIAC PACING, P367
[9]   Cardiac pacing, 1960-1985 - A quarter century of medical and industrial innovation [J].
Jeffrey, K ;
Parsonnet, V .
CIRCULATION, 1998, 97 (19) :1978-1991
[10]   Pacemaker mode selection and survival: a plea to apply the principles of evidence based medicine to cardiac pacing practice [J].
Lamas, GA .
HEART, 1997, 78 (03) :218-220