Minimization of over-ranging in helical volumetric CT via hybrid cone beam image reconstruction - Benefits in dose efficiency

被引:17
作者
Tang, Xiangyang [1 ]
Hsieh, Jiang [1 ]
Dong, Fang [2 ]
Fan, Jiahua [2 ]
Toth, Thomas L. [2 ]
机构
[1] GE Healthcare, Appl Sci Lab, Mol Imaging & Computed Tomog, Milwaukee, WI 53201 USA
[2] GE Healthcare, CT Syst & Image Qual, Mol Imaging & Computed Tomog, Milwaukee, WI 53201 USA
关键词
cone-beam volume CT; helical scan; CT reconstruction; 3D reconstruction; reconstruction algorithms; dose efficiency;
D O I
10.1118/1.2936336
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 [临床医学]; 100207 [影像医学与核医学]; 1009 [特种医学];
摘要
Diagnostic computed tomography (CT) images are usually acquired in both helical and axial scans in the clinical applications using cone beam volumetric CT. In addition to faster patient throughput, a helical scan in volumetric CT can provide better image quality because of the satisfaction of data sufficiency condition, and thus has been performed far more frequently so far in the clinic. However, the first and last images in a helical scan are usually prescribed at the locations that are half helical turn indented from the starting and ending points of the scan. Due to such an indention, the dose efficiency of helical scan deteriorates with increasing detector dimension along z direction. To improve the dose efficiency of helical scan in volumetric CT, a hybrid helical cone beam filtered backprojection (CB-FBP) algorithm is presented here to reconstruct helical images beyond the conventional indented image zone. The hybrid algorithm is a combination of the ray-wise three-dimensional (3D) weighted CB-FBP algorithms that are recently proposed for helical and axial CB image reconstructions. Through the hybridization, the ray-wise 3D weighting becomes dependent on both helical pitch and image slice location. Phantom study shows that the conventional indented image zone in helical scan can be extended substantially by using the hybrid algorithm. Consequently, the dose efficiency of volumetric CT in helical scan can be improved significantly. It is believed that, with increasing detector dimension along z direction in cone beam volumetric CT, the hybrid algorithm will become more attractive in clinical applications. (C) 2008 American Association of Physicists in Medicine.
引用
收藏
页码:3232 / 3238
页数:7
相关论文
共 26 条
[1]
Monte Carlo assessment of computed tomography dose to tissue adjacent to the scanned volume [J].
Boone, JM ;
Cooper, VN ;
Nemzek, WR ;
McGahan, JP ;
Seibert, JA .
MEDICAL PHYSICS, 2000, 27 (10) :2393-2407
[2]
The trouble with CTDI100 [J].
Boone, John M. .
MEDICAL PHYSICS, 2007, 34 (04) :1364-1371
[3]
It is time to retire the computed tomography dose index (CTDI) for CT quality assurance and dose optimization [J].
Brenner, DJ ;
McCollough, CH ;
Orton, CG .
MEDICAL PHYSICS, 2006, 33 (05) :1189-1191
[4]
Effective doses in standard protocols for multi-slice CT scanning [J].
Cohnen, M ;
Poll, LW ;
Puettmann, C ;
Ewen, K ;
Saleh, A ;
Mödder, U .
EUROPEAN RADIOLOGY, 2003, 13 (05) :1148-1153
[5]
COMPUTED-TOMOGRAPHY SCANNING WITH SIMULTANEOUS PATIENT TRANSLATION [J].
CRAWFORD, CR ;
KING, KF .
MEDICAL PHYSICS, 1990, 17 (06) :967-982
[6]
Redundant data and exact helical cone-beam reconstruction [J].
Heuscher, D ;
Brown, K ;
Noo, F .
PHYSICS IN MEDICINE AND BIOLOGY, 2004, 49 (11) :2219-2238
[7]
Survey of conventional and spiral CT doses [J].
Hidajat, N ;
Wolf, M ;
Nunnemann, A ;
Liersch, P ;
Gebauer, B ;
Teichgräber, U ;
Schröder, RJ ;
Felix, R .
RADIOLOGY, 2001, 218 (02) :395-401
[8]
A general approach to the reconstruction of x-ray helical computed tomography [J].
Hsieh, J .
MEDICAL PHYSICS, 1996, 23 (02) :221-229
[9]
Investigation of the slice sensitivity profile for step-and-shoot mode multi-slice computed tomography [J].
Hsieh, J .
MEDICAL PHYSICS, 2001, 28 (04) :491-500
[10]
Multi-slice helical CT: Scan and reconstruction [J].
Hu, H .
MEDICAL PHYSICS, 1999, 26 (01) :5-18