Evaluation of 25-hydroxy vitamin D assay on the immunodiagnostic systems iSYS analyser

被引:27
作者
Cluse, Zee N. [1 ]
Fudge, Andrew N. [1 ]
Whiting, Malcolm J. [1 ]
McWhinney, Brett [2 ]
Parkinson, Ian [3 ]
O'Loughlin, Peter D. [4 ]
机构
[1] SA Pathol, Chem Pathol Directorate, Metab Lab, Adelaide, SA 5000, Australia
[2] HPLC, Dept Chem Pathol, MS Chromatog Sect, LC MS, Brisbane, Qld 4029, Australia
[3] SA Pathol, Surg Pathol Directorate, Bone & Joint Res Lab, Adelaide, SA 5000, Australia
[4] Univ Adelaide, Endocrine Bone Res Lab, Adelaide, SA 5000, Australia
关键词
TANDEM MASS-SPECTROMETRY; D INSUFFICIENCY;
D O I
10.1258/acb.2011.011018
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Background: We evaluated the recently released chemiluminescence assay for 25-hydroxy vitamin D (25-OHD) on the Immunodiagnostic Systems iSYS (IDS-iSYS) automated analyser. Methods: The IDS-iSYS comparison was performed using patient samples previously measured for 25-OHD by a validated liquid chromatography-tandem mass spectrometry (LC-MS/MS) method (n = 119) and an IDS enzyme immunoassy (IDS-EIA) method (n = 64). Limit of detection and limit of quantification were determined from a precision profile. Imprecision was assessed using quality control material and pooled serum. External QAP material (Vitamin D External Quality Assessment Scheme, UK) was analysed to establish inaccuracy. Linearity was assessed by two dilution studies. Cross-reactivity was determined by three serial dilution studies of patient samples with known 25-OHD2 concentrations. Results: The IDS-iSYS correlated well with both established methods (iSYS = 1.03LC-MS/MS - 6.53, R-2 = 0.82 and iSYS = 1.07IDS-EIA - 1.61, R-2 = 0.86). Imprecision of the iSYS assay for IDS control material was 13.4% at 32 nmol/L, 10% at 78 nmol/L, 9.4% at 161 nmol/L, and for the pooled material 9.3% at 72 nmol/L and 5.6% at 158 nmol/L. The evaluation found the assay to be highly accurate (IDS-iSYS = 0.93ALTM + 3.79, R-2 = 0.94) and linear (obs(1) = 0.93exp(1) - 5.05, R-2 = 0.99 (P = 0.256); and obs(2) = 0.97exp(2) + 6.07, R-2 = 0.97 (P = 0.654); ALTM, all-laboratory trimmed mean). Cross-reactivity studies demonstrated no significant difference to the calculated total 25-OHD as measured by LC-MS/MS. Conclusions: Even though the imprecision of the iSYS was found to be greater than that of the LC-MS/MS and EIA methods, the performance characteristics of the IDS-iSYS 25-OHD assay are suitable for routine diagnostic purposes on a high throughput automated analyser.
引用
收藏
页码:159 / 165
页数:7
相关论文
共 19 条
[1]  
[Anonymous], 2006, SOLAR ULTRAVIOLET RA
[2]  
[Anonymous], 2010, IN VITR DIAGN IDS IS
[3]  
Armbruster DA., 2008, CLIN BIOCHEM REV, V29 Suppl 1, pS49, DOI DOI 10.3390/S20236922
[4]   Current status of clinical 25-hydroxyvitamin D measurement: An assessment of between-laboratory agreement [J].
Binkley, Neil ;
Krueger, Diane C. ;
Morgan, Sarah ;
Wiebe, Donald .
CLINICA CHIMICA ACTA, 2010, 411 (23-24) :1976-1982
[5]   Analytical and clinical validation of the 25 OH vitamin D assay for the LIAISON® automated analyzer [J].
Ersfeld, DL ;
Rao, DS ;
Body, JJ ;
Sackrison, JL ;
Miller, AB ;
Parikh, N ;
Eskridge, TL ;
Polinske, A ;
Olson, GT ;
MacFarlane, GD .
CLINICAL BIOCHEMISTRY, 2004, 37 (10) :867-874
[6]   Health benefits of higher serum 25-hydroxyvitamin D levels in The Netherlands [J].
Grant, William B. ;
Schuitemaker, Gert E. .
JOURNAL OF STEROID BIOCHEMISTRY AND MOLECULAR BIOLOGY, 2010, 121 (1-2) :456-458
[7]   Sunlight and vitamin D for bone health and prevention of autoimmune diseases, cancers, and cardiovascular disease [J].
Holick, MF .
AMERICAN JOURNAL OF CLINICAL NUTRITION, 2004, 80 (06) :1678S-1688S
[8]   Vitamin D Status: Measurement, Interpretation, and Clinical Application [J].
Holick, Michael F. .
ANNALS OF EPIDEMIOLOGY, 2009, 19 (02) :73-78
[9]   Relationship between serum 25-hydroxyvitamin D and bone resorption markers in vitamin D insufficiency [J].
Jesudason, D ;
Need, AG ;
Horowitz, M ;
O'Loughlin, PD ;
Morris, HA ;
Nordin, BEC .
BONE, 2002, 31 (05) :626-630
[10]   A simple automated solid-phase extraction procedure for measurement of 25-hydroxyvitamin D3 and D2 by liquid chromatography-tandem mass spectrometry [J].
Knox, Susan ;
Harris, John ;
Calton, Lisa ;
Wallace, A. Michael .
ANNALS OF CLINICAL BIOCHEMISTRY, 2009, 46 :226-230