Acetylsalicylic acid determination in pharmaceutical samples by FIA-potentiometry using a salicylate-sensitive tubular electrode with an ethylene-vinyl acetate membrane

被引:44
作者
Rover, L [1 ]
Garcia, CAB [1 ]
Neto, CD [1 ]
Kubota, LT [1 ]
Galembeck, F [1 ]
机构
[1] UNICAMP, Inst Chem, BR-13083970 Campinas, SP, Brazil
关键词
acetylsalicylic acid determination; ethylene-vinyl acetate; salicylate-sensitive tubular detector; FIA;
D O I
10.1016/S0003-2670(97)00723-X
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The construction of a salicylate-sensitive tubular potentiometric electrode for flow injection analysis (FIA) is described, in which an ethylene-vinyl acetate (EVA) copolymeric membrane, 40% in vinyl acetate, is applied directly onto a conducting support consisting of a mixture of an epoxy resin and graphite. The tubular salicylate ion-selective electrode is based on tricaprylyl-trimethyl-ammonium (Aliquat(TM)) salicylate as the ion-exchanger, supported on the polymeric matrix without plasticizer solvent. The tubular electrode showed a response of 58.3 mV per concentration decade within the 5.0x10(-3)-5.0x10(-1) mol l(-1) salicylate range in 0.1 mol l(-1) Tris-HCl buffer (pH=7.5). Iodide is the main analytical interferent; the electrode response time was 5 s at 25 degrees C. Determinations of acetylsalicylic acid in tablet samples by the proposed method exhibit relative errors lower than 2%, when compared with a spectrophotometric method. The useful lifetime of the tubular sensor is longer than 3 months. (C) 1998 Elsevier Science B.V.
引用
收藏
页码:103 / 109
页数:7
相关论文
共 23 条
[1]  
ALCADA MNM, 1992, J PHARM BIOMED ANAL, V1, P757
[2]   INDOMETHACIN ION-SELECTIVE ELECTRODE BASED ON A BIS(TRIPHENYLPHOSPHORANYLIDENE)AMMONIUM INDOMETHACIN COMPLEX [J].
AUBECK, R ;
BRAUCHLE, C ;
HAMPP, N .
ANALYST, 1991, 116 (08) :811-814
[3]  
CHAMARRO JA, 1993, ANALYST, V118, P1527
[4]   AN EASILY CONSTRUCTED SALICYLATE-ION-SELECTIVE ELECTRODE FOR USE IN THE INSTRUCTIONAL LABORATORY [J].
CREAGER, SE ;
LAWRENCE, KD ;
TIBBETS, CR .
JOURNAL OF CHEMICAL EDUCATION, 1995, 72 (03) :274-276
[5]   SELECTRODE - UNIVERSAL ION-SELECTIVE ELECTRODE .7. VALINOMYCIN-BASED POTASSIUM ELECTRODE WITH NONPOROUS POLYMER MEMBRANE AND SOLID-STATE INNER REFERENCE SYSTEM [J].
FIEDLER, U ;
RUZICKA, J .
ANALYTICA CHIMICA ACTA, 1973, 67 (01) :179-193
[6]   Salicylate selective electrode based on a biomimetic guanidinium ionophore [J].
Hutchins, RS ;
Bansal, P ;
Molina, P ;
Alajarin, M ;
Vidal, A ;
Bachas, LG .
ANALYTICAL CHEMISTRY, 1997, 69 (07) :1273-1278
[7]  
JAMAL AE, 1991, TALANTA, V38, P929
[8]  
JUN S, 1994, J TRACE ELEM ELECT H, V8, P93
[9]   Ion-selective electrode for salicylate assay in blood serum [J].
Katsu, T ;
Mori, Y .
TALANTA, 1996, 43 (05) :755-759
[10]   FURTHER-STUDIES ON THE POTENTIOMETRIC SALICYLATE RESPONSE OF POLYMERIC MEMBRANES DOPED WITH TIN(IV)-TETRAPHENYLPORPHYRINS [J].
KIBBEY, CE ;
PARK, SB ;
DEADWYLER, G ;
MEYERHOFF, ME .
JOURNAL OF ELECTROANALYTICAL CHEMISTRY, 1992, 335 (1-2) :135-149